Radiance-Invariance Evaluation of a Compact LED-Based Source for Imager Radiance Transfer
- Citations
- 0
Abstract
Significance: Fluorescence imaging remains largely qualitative and device-specific, limiting reproducibility and intersystem comparisons. SI-traceable radiometric characterization requires calibrated sources suitable for transferring known radiance to imaging systems. Aim: Establish and evaluate a compact, calibrated solid-state radiometric emitter target (RET) for SI-traceable radiance transfer, and assess methods for evaluating radiance invariance, imager responsivity, and extension of radiance transfer to digitally defined regions of interest (ROIs). Approach: The radiance invariance of an LED-based RET was evaluated by measuring its radiance across varying source-to-aperture distances and collection apertures. A previously described radiance-transfer formalism based on calibrated source radiance and collection geometry was implemented to determine imager responsivity (R{lambda}), which was evaluated across varying working distances and lens apertures. Grubbs and MAD outlier tests and one-way ANOVA were used to assess radiance and responsivity invariance. Finally, responsivity obtained using digitally defined ROIs was compared with matched physical apertures to assess ROI-based radiance transfer. Results: The measured RET radiance remained invariant across the tested source-to-aperture distances and collection apertures, with no significant differences observed across distance (p = 0.88) or aperture (p = 0.15). Imager responsivity remained stable across working distances for both tested lens configurations (p = 0.48 and p > 0.99). Responsivity was also invariant across lens apertures, except for deviations at f/11 that were consistent with practical aperture tolerances. Digitally defined ROIs reproduced responsivity values obtained using matched physical apertures, demonstrating that the radiance-transfer approach can be extended to image-defined regions. Conclusions: The presented methods provide an approach for evaluating compact LED-based sources for invariant-radiance emission and suitability for SI-traceable radiance transfer. The characterized RET supported stable imager responsivity across changes in imaging geometry and extension to digitally defined ROIs. These methods can be adapted to other LED-based sources, with further angular characterization and uncertainty analysis needed to establish broader applicability.
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Quantitative Fluorescence Imaging with SI-Traceable Radiance: Radiometric Fluorescence Characterization with a Calibrated Solid-State Emitter
Alberto J. Ruiz, Edwin A. Robledo, Eammon A. Littler·bioRxiv·Oct 15, 2025
- Computational Luminance Constancy from Naturalistic Images
Vijay Singh, Nicolas P. Cottaris, … Johannes Burge·Journal of VisionPeer-reviewed·Dec 28, 2018
- Invariant categorical color regions across illuminant change coincide with focal colors
Takuma Morimoto, Yasuki Yamauchi, Keiji Uchikawa·Journal of VisionPeer-reviewed·Feb 13, 2023
- The effect of illumination cues on color constancy in simultaneous identification of illumination and reflectance changes
Lari S. Virtanen, Maria Olkkonen, Toni Saarela·Journal of VisionPeer-reviewed·May 6, 2025
- Characterization of human lightness discrimination thresholds for independent spectral variations
D. J. Reynolds, V. Singh·bioRxiv·Jun 16, 2023
- When the brightest is not the best: illuminant estimation from the geometry of specular highlights
Takuma Morimoto, Robert J. Lee, Hannah E. Smithson·Proceedings of the Royal Society B Biological SciencesPeer-reviewed·Aug 5, 2026