PLoS ONEMolecular BiologyPeer-reviewed journal article
Validation of saline, PBS and a locally produced VTM at varying storage conditions to detect the SARS-CoV-2 virus by qRT-PCR
Caroline NgetsaiD, Victor OsotiiD, Dorcas OkandaiD, Faith Marura, Krupali ShahiD, Henry KaranjaiD, Daisy MugoiD, John N. GitongaiD, Martin N. Mutunga, Clement Lewa, Benedict OrindiiD, Philip BejoniD, Lynette Isabella Ochola‐OyieriD
- Citations
- 12
- FWCI
- 2.47
- Field percentile
- top 12%
Abstract
Coronavirus Disease-2019 tests require a Nasopharyngeal (NP) and/or Oropharyngeal (OP) specimen from the upper airway, from which virus RNA is extracted and detected through quantitative reverse transcription-Polymerase Chain Reaction (qRT-PCR). The viability of the virus is maintained after collection by storing the NP/OP swabs in Viral Transport Media (VTM). We evaluated the performance of four transport media: locally manufactured ("REVITAL") Viral Transport Media (RVTM), Standard Universal Transport Media (SUTM), PBS and 0.9% (w/v) NaCl (normal saline). We used laboratory cultured virus to evaluate: i) viral recovery and maintaining integrity at different time periods and temperatures; ii) stability in yielding detectable RNA consistently for all time points and conditions; and iii) their overall accuracy. Four vials of SARS-CoV-2 cultured virus (2 high and 2 low concentration samples) and 1 negative control sample were prepared for each media type (SUTM, RVTM, PBS and normal saline) and stored at the following temperatures, -80{degrees}C, 4{degrees}C, room temperature (25{degrees}C) and 37{degrees}C for 7 days. Viral Ribonucleic acid (RNA) extractions and qRT-PCR were done on the following days after inoculation with the cultured virus, days 1, 2, 3, 4 and 7 to assess virus stability and viral recovery. CT values fell over time at room temperature, but normal saline, PBS, RVTM and SUTM all showed comparable performance in maintaining virus integrity and stability allowing for the detection of SARS-CoV-2 viral RNA. Overall, this study demonstrated that normal saline, PBS and the locally manufactured VTM can be used for COVID-19 sample collection and testing, thus expanding the range of SARS-CoV-2 viral collection media.
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Dry Swab Method of sample collection for SARS-CoV2 testing can be used for culturing virus
S. Ram, M. G. Enayathullah, … B. Kiran Kumar·bioRxiv·Mar 23, 2021
- Influence of viral transport media and freeze-thaw cycling on the sensitivity of qRT-PCR detection of SARS-CoV-2 nucleic acids
Cian Holohan, Sophia Hanrahan, … Gil U. Lee·NanoscalePeer-reviewed·Jan 1, 2021
- Optimization of SARS-CoV-2 detection by RT-QPCR without RNA extraction
Natacha Mérindol, Geneviève Pépin, … Alexis Danylo·bioRxiv·Apr 6, 2020
- COVID-19 PCR Test Performance For Samples Stored At Ambient Temperature
N. B. Agaoglu, J. Yıldız, … G. Dinler-Doganay·bioRxiv·Jun 15, 2020
- The impact of viral transport media on PCR assay results for the detection of nucleic acid from SARS-CoV-2 and other viruses
P. D. Kirkland, M. J. Frost·bioRxiv·Jun 9, 2020
- A simple, safe and sensitive method for SARS-CoV-2 inactivation and RNA extraction for RT-qPCR
L. Kalnina, A. Mateu-Regue, … A. K. N. Iversen·bioRxiv·Jun 29, 2020