bioRxivNeurosciencePreprint
Use of non-invasive intracranial pressure pulse waveform to monitor patients with End-Stage Renal Disease (ESRD)
Cristiane RickliiD, Laís Daiene CosmoskiiD, Fábio André dos SantosiD, Gustavo Henrique FrigieriiD, Nícollas Nunes RabeloiD, Adriana Fátima Menegat SchuinskiiD, Sérgio MascarenhasiD, José Carlos Rebuglio VellosaiD
Abstract
End-stage renal disease (ESRD) is treated mainly by hemodialysis, however, hemodialysis is associated with frequent complications, some of them involve the increased intracranial pressure. In this context, monitoring the intracranial pressure of these patients may lead to a better understanding of how intracranial pressure morphology varies with hemodialysis. This study aimed to follow-up patients with ESRD by monitoring intracranial pressure before and after hemodialysis sessions using a noninvasive method. We followed-up 42 patients with ESRD in hemodialysis, for six months. Noninvasive intracranial pressure monitoring data were obtained through analysis of intracranial pressure waveform morphology, this information was uploaded to Brain4care(R) cloud algorithm for analysis. The cloud automatically sends a report containing intracranial pressure parameters. In total, 4881 data points were collected during the six months of follow-up. The intracranial pressure parameters (time to peak and P2/P1 ratio) were significantly higher in predialysis when compared to postdialysis for the three weekly sessions and throughout the follow-up period (p<0.01) data showed general improvement in brain compliance after the hemodialysis session. Furthermore, intracranial pressure parameters were significantly higher in the first weekly hemodialysis session (p<0.05). In conclusion, there were significant differences between pre and postdialysis intracranial pressure in patients with ESRD on hemodialysis. Additionally, the pattern of the intracranial pressure alterations was consistent over time suggesting that hemodialysis can improve time to peak and P2/P1 ratio which may reflect in brain compliance.
- Citations
- 0
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Nonlinear Parameter and State Estimation Approach in End-stage Kidney Disease Patients
R. Abohtyra, T. L. Vincent·bioRxiv·Apr 2, 2022
- Nonlinear Parameter and State Estimation Approach for Intradialytic Measurement of Absolute Blood Volume
R. Abohtyra, T. Vincent, D. Schneditz·bioRxiv·Aug 12, 2023
- Real-time non-invasive intracranial state estimation using unscented Kalman filter
Chanki Park, Seungjun Ryu, … Boreom Lee·IEEE Transactions on Neural Systems and Rehabilitation EngineeringPeer-reviewed·Jul 31, 2019
- Cascaded Nano-ring Dialyzer for Continuous Regenerative Peritoneal Dialysis
Wonseok Kim, Kihong Kim, … Sung Jae Kim·bioRxiv·Sep 21, 2026
- Noninvasive optical monitoring of cerebral hemodynamics in a preclinical model of neonatal intraventricular hemorrhage
Jyoti V. Jethe, YuBing Y. Shen, … Jonathan A. N. Fisher·Frontiers in PediatricsPeer-reviewed·Mar 10, 2025
- Body Fluid Estimation during Standard Ultrafiltration in Chronic Kidney Disease