Multiple Channel Electrocardiogram QRS Detection by Temporal Pattern Search
Abstract
A highly constrained temporal pattern search ("TEPS") based multiple channel heartbeat detector is described. TEPS generates sequences of peaks and statistically scores them according to: 1) peak time coherence across channels; 2) peak prominence; 3) temporal regularity; and 4) number of skipped beats. TEPS was tested on 31 records of three channel capacitive electrode data from the UnoViS automobile database. TEPS showed a sensitivity (SE) of 91.3% and a false discovery rate (FDR) of 3.0% compared to an SE and FDR of 75.3% and 65.0% respectively for a conventional single channel detector (OSEA) applied separately to the three channels. The peak matching window was 30ms. The percentage of 5 second segments with average heart rates within 5 beats/minute of reference was also measured. In 6 of the 31 records, TEPS percentage was at least 30% greater than OSEAs. TEPS was also applied to synthetic data comprising a known signal corrupted with calibrated amounts of noise. At a fixed SE of 85%, increasing the number of channels from one to two resulted in an improvement of approximately 5dB in noise resistance, while increasing the number of channels from two to four resulted in an improvement of approximately 3dB in noise resistance. The quantification of noise resistance as a function of the number of channels could help guide the development of wearable electrocardiogram monitors.
02 /Specification & Classification
Click any tag to browse matching papersReproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- QRS Detection by Rules Based Multiple Channel Combinatorial Optimization
B. Hopenfeld·bioRxiv·Jun 16, 2023
- QRS Detection by Combinatorial Optimization With MLP Assisted Peak Scoring
Bruce Hopenfeld·bioRxiv·Apr 22, 2026
- Fetal QRS Detection by Multiple Channel Temporal Pattern Search Applied to the NInFEA Database
B. R. Hopenfeld·bioRxiv·Aug 29, 2022
- Segment Based Pattern Analysis Reveals a Persistent Regular Rhythm in the Motion Artifact Record of the MIT-BIH Noise Stress Test Database
B. R. Hopenfeld·bioRxiv·Oct 18, 2022
- A new QRS detector stress test combining temporal jitter and F-score (JF) reveals significant performance differences amongst popular detectors
Bernd Porr, Peter W. Macfarlane·PLoS ONEPeer-reviewed·Nov 11, 2024
- Peak Space Motion Artifact Cancellation Applied to Textile Electrode Waist Electrocardiograms Recording During Outdoors Walking and Jogging
B. R. Hopenfeld·bioRxiv·Jan 7, 2022