Sound comparison of seven TMS coils at matched stimulation strength

Lari M. KoponeniD, Stefan M. GoetziD, Debara Lyn TucciiD, Angel V. PetercheviD

PDFDOIPreprint

Abstract

BackgroundAccurate data on the sound emitted by transcranial magnetic stimulation (TMS) coils is lacking.\n\nMethodsWe recorded the sound waveforms of seven coils with high bandwidth. We estimated the neural stimulation strength by measuring the induced electric field and applying a strength-duration model to account for different waveforms.\n\nResultsAcross coils, at maximum stimulator output and 25 cm distance, the sound pressure level (SPL) was 98-125 dB(Z) per pulse and 75-97 dB(A) for a 15 Hz pulse train. At 5 cm distance, these values were estimated to increase to 112-139 dB(Z) and 89-111 dB(A), respectively.\n\nConclusionsThe coils sound was below, but near, relevant exposure limits for operators and may exceed some limits for the subject. Exposure standards may inadequately capture some risks to hearing. For persons near operating TMS coils we recommend hearing protection, and we consider it essential for the TMS subject.\n\nHighlights O_LICoil click varies by over 20 dB(Z) between TMS coils at matched stimulation strength.\nC_LIO_LIClose to TMS coil, sound pressure level may reach nearly 140 dB(Z).\nC_LIO_LIFor rTMS, the continuous sound level can exceed 110 dB(A).\nC_LIO_LIHearing protection is recommended during TMS, especially for the subject.\nC_LI

Citations
45
FWCI
3.00
Field percentile
top 8%

Reproducibility ledger

Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.

Sign in to file a reproducibility report.