bioRxivBiophysicsPreprint
SCKAnalyzer: A Web Tool for Single-Cycle Kinetics Analysis and Simulation
Florian MalardiD, Jean-Michel BlanciD, Eric Roubin, Thomas SchäferiD, Carmelo Di PrimoiD
Abstract
Surface plasmon resonance (SPR) and biolayer interferometry (BLI) are well-established label-free analytical technologies used to determine biomolecular interaction rates (ka, kd) and equilibrium constants (KD). While assays historically rely on multi-cycle kinetics (MCK), this approach requires intermediate surface regeneration, risking sensor degradation and increasing reagent consumption. Single-cycle kinetics (SCK) avoids regeneration by sequentially injecting increasing analyte concentrations. However, implementing SCK workflows on biosensor platforms like the OctetRR8 BLI instrument (Sartorius) remains a challenge due to the lack of built-in or online data analysis tools. To address this, we introduce SCKAnalyzer, a free, user-friendly web application designed for processing and fitting SCK datasets. SCKAnalyzer accepts column-based text files and native raw BLI data (.frd format) from Sartorius instruments, performing data pre-processing and non-linear fitting to a 1:1 interaction model. Additionally, it offers advanced simulation capabilities that allow variable association and dissociation durations across sequential injection steps, a feature absent in existing platforms such as BiacoreTMSimul8. The performance of SCKAnalyzer was experimentally validated using a 3-step SCK method (SCK3) for DNA/RNA duplex formation on a Biacore T200, as well as 4-step (SCK4) and 5-step (SCK5) methods for DNA/DNA duplex formation on an OctetRR8 BLI instrument. Binding parameters (ka, kd, KD) derived from SCKAnalyzer showed excellent agreement with values from established proprietary softwares such as Biacore T200 Evaluation Software v3.2.1 and BIAevaluation v4.1.1. SCKAnalyzer provides a practical alternative to MCK, reducing probe consumption and extending SCK utility across biosensor platforms. SCKAnalyzer is available at sck.iecb.u-bordeaux.fr (source code: github.com/synthaze/sckanalyzer).
- research focus: benchmarking
- research focus: software / tool
- research focus: database / resource
- approach: statistical model
- Citations
- 0
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- The SpyBLI cell-free pipeline for the rapid quantification of binding kinetics from crude samples
Olga Predeina, Misha Atkinson, … Pietro Sormanni·RSC Chemical BiologyPeer-reviewed·Jan 1, 2025
- Digital molecular kinetics enabled by large-scale nanopore arrays for multidimensional biophysical profiling
Kaiju Li, Yan Wang, … Yaozhong Zou·bioRxiv·Oct 29, 2026
- waveRAPID(R) - a robust assay for high-throughput kinetic screens with the Creoptix(R) WAVEsystem
O. Kartal, F. Andres, … K. Cottier·bioRxiv·Feb 5, 2021
- Novel sensor-integrated proteome on chip (SPOC) platform with thousands of folded proteins on a 1.5 sq-cm biosensor chip to enable high-throughput real-time label-free screening for kinetic analysis.
Chidozie Victor Agu, Rebecca L. Cook, … Bharath R. Takulapalli·Communications BiologyPeer-reviewed·Mar 21, 2025
- Surface Transmon Resonance (STR): a handheld nanogap biosensor for real-time, label-free molecular binding kinetics
Bryan Karekin Chantigian, Sang‐Hyun Oh·npj BiosensingPeer-reviewed·Feb 19, 2026
- Continuous Titration Based Method For Rapid In-solution Analysis Of Non-covalent Interactions
Philipp Willmer, Adam C. Hundahl, … Henrik M. Jensen·Chemistry - MethodsPeer-reviewed·Feb 14, 2025