Journal of Applied PhysicsNeurosciencePeer-reviewed journal article
Magnetoelectrics Enables Large Power Delivery to mm-Sized Wireless Bioelectronics
Wonjune KimiD, C. Anne TuppeniD, Fatima T. AlrashdaniD, Amanda L. SingeriD, Rachel WeirnickiD, Jacob T. RobinsoniD
Abstract
To maximize the capabilities of minimally invasive implantable bioelectronic devices, we must deliver large amounts of power to small implants; however, as devices are made smaller, it becomes more difficult to transfer large amounts of power without a wired connection. Indeed, recent work has explored creative wireless power transfer (WPT) approaches to maximize power density (the amount of power transferred divided by receiver footprint area (length x width)). Here, we analyzed a model for WPT using magnetoelectric (ME) materials that convert an alternating magnetic field into an alternating voltage. With this model, we identify the parameters that impact WPT efficiency and optimize the power density. We find that improvements in adhesion between the laminated ME layers, clamping, and selection of material thicknesses lead to a power density of 3.1 mW/mm2, which is over 4 times larger than previously reported for mm-sized wireless bioelectronic implants at a depth of 1 cm or more in tissue. This improved power density allows us to deliver 31 mW and 56 mW to 10-mm2 and 27-mm2 ME receivers, respectively. This total power delivery is over 5 times larger than similarly sized bioelectronic devices powered by radiofrequency electromagnetic waves, inductive coupling, ultrasound, light, capacitive coupling, or previously reported magnetoelectrics. This increased power density opens the door to more power-intensive bioelectronic applications that have previously been inaccessible using mm-sized battery-free devices.
- Citations
- 31
- FWCI
- 2.31
- Field percentile
- top 11%
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Magnetoelectric materials for miniature, wireless neural stimulation at therapeutic frequencies
A. Singer, S. Dutta, … J. T. Robinson·bioRxiv·Mar 15, 2020
- Distributed resonant coupling enables high efficiency power transfer to mm-scale bioelectronics
Sunghoon Rho, Garrett Knuf, … Siddharth R. Krishnan·bioRxiv·Jul 17, 2026
- Powering electronic implants by high frequency volume conduction: in human validation
J. Minguillon, M. Tudela-Pi, … A. Ivorra·bioRxiv·Mar 15, 2021
- Ultra-compact Dual-band Smart NEMS Magnetoelectric Antennas for Simultaneous Wireless Energy Harvesting and Magnetic Field Sensing
M. Zaeimbashi, M. Nasrollahpour, … N. Sun·bioRxiv·Jun 22, 2020
- Autonomous Wireless System for Robust and Efficient Inductive Power Transmission to Multi-Node Implants
P. Feng, T. Constandinou·bioRxiv·Feb 1, 2021
- Floating EMG Sensors and Stimulators Wirelessly Powered and Operated by Volume Conduction for Networked Neuroprosthetics
L. Becerra-Fajardo, M. O. Krob, … A. Ivorra·bioRxiv·Mar 9, 2022