Journal of Neural EngineeringNeurosciencePeer-reviewed journal article
Accuracy of robotic coil positioning during transcranial magnetic stimulation
Stefan M. GoetziD, I. Cassie Kozyrkov, Bruce M. LuberiD, Sarah Hollingsworth LisanbyiD, David L. K. MurphyiD, Warren M. GrilliD, Angel V. PetercheviD
Abstract
ObjectiveRobotic positioning systems for transcranial magnetic stimulation (TMS) promise improved accuracy and stability of coil placement, but there is limited data on their performance. This text investigates the usability, accuracy, and limitations of robotic coil placement with a commercial system, ANT Neuro, in a TMS study.\n\nApproach21 subjects underwent a total of 79 TMS sessions corresponding to 160 hours under robotic coil control. Coil position and orientation were monitored concurrently through an additional neuronavigation system.\n\nMain ResultsRobot setup took on average 14.5 min. The robot achieved low position and orientation error with median 1.34 mm and 3.48{degrees}. The error increased over time at a rate of 0.4%/minute for both position and orientation.\n\nSignificanceAfter the elimination of several limitations, robotic TMS systems promise to substantially improve the accuracy and stability of manual coil position and orientation. Lack of pressure feedback and of manual adjustment of all coil degrees of freedom were limitations of this robotic system.
- Citations
- 43
- FWCI
- 1.87
- Field percentile
- top 15%
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Characterizing an electronic-robotic targeting platform for precise and fast brain stimulation with multi-locus transcranial magnetic stimulation
Renan H. Matsuda, Victor Hugo Souza, … Oswaldo Baffa·Physics in Medicine and BiologyPeer-reviewed·Jul 23, 2025
- Inside-out tracking and projection mapping for robot-assisted transcranial magnetic stimulation
- Quantitative Impact of Coil Misalignment and Misplacement in Transcranial Magnetic Stimulation
Max Koehler, Thomas Kammer, Stefan M. Goetz·bioRxiv·Nov 18, 2023
- TAP: Targeting and analysis pipeline for optimization and verification of coil placement in transcranial magnetic stimulation
Moritz Dannhauer, Ziping Huang, … Angel V. Peterchev·Journal of Neural EngineeringPeer-reviewed·Apr 1, 2022
- MRI-free Virtual Neuronavigation for TMS
Yishai Valter, Dennis Q. Truong, … Abhishek Datta·bioRxiv·Apr 10, 2025
- EEG-guided robotic control of transcranial magnetic stimulation
Renan H. Matsuda, Matilda Makkonen, … Pantelis Lioumis·bioRxiv·May 19, 2026