RF Channel-Based Adaptive Beamforming for 3D Ultrasound Localization Microscopy
Georges ChabouhiD, Baptiste PialotiD, Louise DenisiD, Raphaël DumasiD, Olivier CoutureiD, Pauline Muleki-SeyaiD, François VarrayiD
Abstract
Ultrasound Localization Microscopy (ULM) has been applied in various preclinical settings and in the clinic to reveal the microvasculature in deep organs. However, most ULM images employ standard Delay-and-Sum (DAS) beamforming. In standard ULM conditions, lengthy acquisition times are required to fully reconstruct small vessels due to the need for spatially isolated microbubbles, resulting in low temporal resolution. When microbubbles are densely packed, localizing a point spread function with significant main and side lobes becomes challenging due to matrix arrays low signal-to-noise ratio and spatial resolution. In this work, we applied adaptive beamforming such as high order DAS known as (pDAS), Coherence Factor (CF), Coherence Factor with Gaussian Filtering (CFGF), and statistical interpretation of beamforming (iMAP) to provide a more complete 3D ULM maps in vitro and in vivo (rat kidney). Specifically, the CF and 1MAP adaptive beamformers achieved higher resolution (32.9 microns and 27.2 microns respectively), as measured by the Fourier Shell Correlation (FSC), compared to the standard DAS beamformer, which had an FSC value of 38.6 microns.
- organism: rat
- assay: imaging
- research focus: imputation
- approach: deep learning
- approach: statistical model
- Citations
- 2
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- An Anatomically and Hemodynamically Realistic Simulation Framework for 3D Ultrasound Localization Microscopy
H. Belgharbi, J. Poree, … L. Milecki·bioRxiv·Oct 8, 2021
- Backscattering Amplitude in Ultrasound Localization Microscopy
N. Renaudin, S. Pezet, … M. Tanter·bioRxiv·Apr 27, 2023
- Microbubble Backscattering Intensity Improves the Sensitivity of Three-dimensional (3D) Functional Ultrasound Localization Microscopy (fULM)
YiRang Shin, Qi You, … Pengfei Song·IEEE Transactions on Medical ImagingPeer-reviewed·Feb 3, 2026
- Context-Aware Deep Learning Enables High-Efficacy Localization of High Concentration Microbubbles for Super-Resolution Ultrasound Localization Microscopy
Y. Shin, M. Lowerison, … P. Song·bioRxiv·Apr 21, 2023
- Nanobubble-Based Ultrasound Localization Microscopy through Interactive Adaptive Processing
Grigori Shapiro, Yarin Gershman, … Tali Ilovitsh·bioRxiv·Jul 17, 2026
- Deep Learning-based Microbubble Localization for Ultrasound Localization Microscopy
X. Chen, M. Lowerison, … P. Song·bioRxiv·Feb 2, 2022