Multiplane and Spectrally-Resolved Single Molecule Localization Microscopy with Industrial Grade CMOS Cameras

H. Babcock

PDFDOIPreprint

Abstract

In this work we explore the use of industrial grade CMOS cameras for single molecule localization microscopy (SMLM). We show that the performance of these cameras in single imaging plane SMLM applications is comparable to much more expensive scientific CMOS (sCMOS) cameras. We show that these cameras can be used in more demanding biplane, multiplane and spectrally resolved SMLM applications. The 10-40x reduction in camera cost makes it practical to build SMLM setups with 4 or more cameras. In addition we provide open-source software for simultaneously controlling multiple CMOS cameras and for the reduction of the movies that are acquired to super-resolution images.