Quantification of thermal side effects during pulsed field ablation
Marcela Mercado‐MontoyaiD, Tatiana Gomez BustamanteiD, Steven Richard MickelseniD, Erik Benjamin KulstadiD, Ana González‐SuáreziD, Lawrence OverzetiD
Abstract
Background and AimsPulsed field ablation (PFA) has been described as non-thermal, but abundant data exist in oncology applications, and growing data are emerging in cardiology, highlighting that thermal effects are in fact present with PFA. Our objective was to develop a reliable model of the thermal effects arising from PFA of myocardial and esophageal tissue over a range of typical peak voltage operating conditions. MethodsWe developed a three-dimensional computer model of the left atrium that can quantify the thermal effects from PFA applications. Energy was applied using a bipolar configuration, and far-field and symmetry boundaries were set as electrically insulating. The model utilizes a monophasic waveform with a pulse width of 100 s and gap between pulses of 1 s was applied for a total of 50 pulses in a single train, with variations to this easily implemented to accommodate various conditions. ResultsOver a range of peak voltage operating conditions (1 kV, 1.5 kV, and 2 kV), minimal temperature rise in the esophagus was seen with 1 kV pulses (corresponding to 215 J input), but with 1.5 kV and 2 kV peak voltages (corresponding to 570 J and 1.23 kJ), temperature elevations reaching 46.3{degrees}C and > 62 {degrees}C were seen, respectively. These elevations occurred after only a single pulse train of 50 pulses, implying that further elevations in temperature would be seen with subsequent applications. These findings concur with data published in other fields of medicine where pulsed field treatments are utilized. ConclusionsThermal effects from PFA occur and can be quantified with in-silico modeling. The model described here offers an efficient means of determining temperature effects from PFA over various operating conditions. Energy levels used clinically appear to have the potential to induce collateral thermal injury with repeated applications of pulsed field energy.
- Citations
- 2
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Comparison of therapeutic effects between pulsed field ablation and cryoballoon ablation in the treatment of atrial fibrillation: a systematic review and me-ta-analysis
Yun Wan, Shuting Zeng, … Jun Luo·Cardiology in ReviewPeer-reviewed·Nov 14, 2024
- Bioimpedance-assisted characterization of cardiac electroporation and anisotropic homogenization by pulsed field ablation
Edward J. Jacobs, Pedro P Santos, … Rafael V. Davalos·bioRxiv·Mar 20, 2026
- Endoluminal catheter pulsed field ablation for the treatment of atherosclerotic vascular disease
Zaid S. Salameh, Edward J. Jacobs, Rafael V. Davalos·Annals of Biomedical EngineeringPeer-reviewed·Feb 28, 2026
- Dual cryo-radiofrequency ablation enhances lesion depth in beating human left ventricular preparations
Matthew S. Sulkin, Jacob I. Laughner, … Igor R. Efimov·bioRxiv·Nov 14, 2018
- Computer Modeling of Radiofrequency Cardiac Ablation including Heartbeat-Induced Electrode Displacement
J. J. Perez, A. Gonzalez-Suarez, … E. Berjano·bioRxiv·Dec 11, 2021
- Study of flow effects on temperature-controlled radio-frequency ablation using phantom experiments and forward simulations