bioRxivBioengineering
Automated assembly of high-density carbon fiber electrode arrays for single unit electrophysiological recordings
T. Dong, L. Chen, P. R. Patel, J. M. Richie, C. A. Chestek, A. J. Shih
Abstract
ObjectiveCarbon fiber (CF) is good for chronic neural recording due to the small diameter (7 {micro}m), high Youngs modulus, and low electrical resistance, but most high-density carbon fiber (HDCF) arrays are manually assembled with labor-intensive procedures and limited by the accuracy and repeatability of the operator handling. A machine to automate the assembly is desired. ApproachThe HDCF array assembly machine contains: 1) a roller-based CF extruder, 2) a motion system with three linear and one rotary stages, 3) an imaging system with two digital microscope cameras, and 4) a laser cutter. The roller-based extruder automatically feeds single CF as raw material. The motion system aligns the CF with the array backend then places it. The imaging system observes the relative position between the CF and the backend. The laser cutter cuts off the CF. Two image processing algorithms are implemented to align the CF with the support shanks and circuit connection pads. Main resultsThe machine was capable of precisely handling 6.8 m carbon fiber electrodes. Each electrode was placed into a 12 m wide trenches in a silicon support shank. Two HDCF arrays with 16 CF electrodes populated on 3 mm shanks (with 80 m pitch) were fully assembled. Impedance measurements were found to be in good agreement with manual assembled arrays. One HDCF array was implanted in the motor cortex in an anesthetized rat and was able to detect single unit activity. SignificanceThis machine can eliminate the manual labor-intensive handling, alignment and placement of single carbon fiber during assembly, providing a proof-of-concepts towards fully automated HDCF array assembly and batch production.
- organism: rat
- assay: electrophysiology
- research focus: cell type annotation
- approach: experimental assay
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- A high-density carbon fiber neural recording array technology
T. L. Massey, S. R. Santacruz, … M. M. Maharbiz·bioRxiv·May 13, 2018
- Benchtop Carbon Fiber Microelectrode Array Fabrication Toolkit
J. M. Richie, P. R. Patel, … C. A. Chestek·bioRxiv·Mar 22, 2021
- Sharpened and mechanically durable carbon fiber electrode arrays for neural interfacing
E. J. Welle, J. E. Woods, … C. A. Chestek·bioRxiv·Jan 21, 2021
- 64-Channel Carbon Fiber Electrode Arrays for Chronic Electrophysiology
Grigori Guitchounts, David Cox·Scientific ReportsPeer-reviewed·Mar 2, 2020
- Carbon Fiber On Polyimide Ultra-Microelectrodes
Winthrop F. Gillis, Charles A. Lissandrello, … Timothy J. Gardner·Journal of Neural EngineeringPeer-reviewed·Sep 14, 2017
- Carbon Fiber Electrodes for Intracellular Recording and Stimulation
Yu Huan, Jeffrey P. Gill, … Hillel J. Chiel·Journal of Neural EngineeringPeer-reviewed·Nov 26, 2021