A metal artifact reduction method for small field of view CT imaging.
Abstract
To reduce metal artifacts, several sinogram inpainting-based metal artifact reduction (MAR) methods have been proposed where projection data within the metal trace region of the sinogram are treated as missing and subsequently estimated. However, these methods generally assume data truncation does not occur and all metal objects reside inside the field-of-view (FOV). For small FOV imaging, these assumptions are violated, and thus existing inpainting-based MAR methods would not be effective. In this paper, we propose a new MAR method to reduce metal artifacts effectively in the presence of data truncation. The main idea behind the proposed method is the synthesis of a sinogram, which is treated as the originally measured sinogram. First, an initial reconstruction step is performed to remove truncation artifacts. The next step consists of a forward projection of the small FOV image, replacing the original sinogram with the synthesized sinogram. The final step is the application of sinogram inpainting based MAR methods using the synthesized sinogram. Verification of the proposed method was performed for three situations: extended cardiac-torso (XCAT) simulation, clinical data, and experimental data. The proposed method was applied with linear MAR (LMAR) and normalized MAR (NMAR), and the performance of the proposed method was compared with that of the previous method. For quantitative evaluation, normalized mean squared error (NMSE) and structure similarity (SSIM) were used. The results show the effectiveness of the proposed method to reduce residual metal artifacts, which are present in the results obtained with the previous method. The evaluation results using NMSE and SSIM also indicate that the proposed method is more effective than the previous method.
02 /Specification & Classification
Click any tag to browse matching papers- Assay
- Approach
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- QSMxT: Robust Masking and Artefact Reduction for Quantitative Susceptibility Mapping
Ashley Wilton Stewart, Simon Robinson, … Steffen Bollmann·Magnetic Resonance in MedicinePeer-reviewed·Oct 22, 2021
- Sinogram Domain Angular Upsampling of Sparse-View Micro-CT with Dense Residual Hierarchical Transformer and Noise-Aware Loss
A. S. Adishesha, D. J. Vanselow, … S. X. Huang·bioRxiv·May 9, 2023
- Evolution from total variation to nonlinear sparsifying transform for sparse-view CT image reconstruction
Jian Wen Dong, Chunxiao Han, … Yanqiu Che·bioRxiv·Sep 27, 2019
- Data-driven system matrix manipulation enabling fast functional imaging and intra-image nonrigid motion correction in tomography
Peng Hu, Xin Tong, … Lihong V. Wang·IEEE Transactions on Medical ImagingPeer-reviewed·Sep 22, 2025
- Tiled Reconstruction Improves Structured Illumination Microscopy
D. P. Hoffman, E. Betzig·bioRxiv·Jan 6, 2020
- A Modified Tseng Algorithm Approach to Restoring Thoracic Diseases' Computerized Tomography Images
Huzaifa Umar, Abubakar Adamu, … Dilber Uzun Ozsahin·bioRxiv·Dec 27, 2023