Trueness of digital intraoral impression in reproducing multiple implant position
Abstract
The aim of this study was to evaluate the trueness of 5 intraoral scanners (IOSs) for digital impression of simulated implant scan bodies in a complete-arch model. A 3D printed full-arch mandible model made of Co-Cr with a total of 6 bilaterally positioned cylinders in the canine, second premolar, and second molar area served as the study model. Digital scans of the model were made with a reference scanner (steroSCAN neo) and 5 IOSs (CEREC Omnicam, CS3600, i500, iTero Element, and TRIOS 3) (n=10). For each IOSs dataset, the XYZ coordinates of the cylinders were obtained from the reference point and the deviations from the reference scanner were calculated using a 3D reverse engineering program (Rapidform). The trueness values were analyzed by Kruskal-Wallis test and Mann-Whitney post hoc test. Direction and amount of deviation differed among cylinder position and among IOSs. Regardless of the IOS type, the cylinders positioned on the left second molar, nearest to the scanning start point, showed the smallest deviation. The deviation generally increased further away from scanning start point towards the right second molar. TRIOS 3 and i500 outperformed the other IOSs for complete-arch digital impression. The accuracy of the CEREC Omnicam, CS3600, and iTero Element were similar on the left side, but they showed more deviations on the right side of the arch when compared to the other IOSs. The accuracy of IOS is still an area that needs to be improved.
02 /Specification & Classification
Click any tag to browse matching papers- Assay
- Research Focus
- Approach
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Development of intra-oral automated landmark recognition (ALR) for dental and occlusal outcome measurements
Brénainn Woodsend, Eirini Koufoudaki, … Peter Anthony Mossey·European Journal of OrthodonticsPeer-reviewed·Mar 15, 2021
- Correlation of two different devices for the evaluation of primary implant stability depending on dental implant length and bone density: an in vitro study
- Why detailed modelling matters in the pre-clinical evaluation of temporomandibular joint implants
Girish Chandra, Rajdeep Ghosh, … Kaushik Mukherjee·bioRxiv·Apr 28, 2025
- Assessing tooth wear progression in primates: A longitudinal study using intraoral scanning technology
Ian Towle, Kristin L. Krueger, … Leslea J. Hlusko·PeerJPeer-reviewed·Jul 9, 2024
- Biomechanical Evaluation of Temporomandibular Joint Implants and Periprosthetic Bone under Unilateral and Bilateral Clenching
Rajdeep Ghosh, Girish Chandra, … Kaushik Mukherjee·Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in MedicinePeer-reviewed·May 27, 2025
- Modifiable clinical dental impression methods to obtain whole-mouth and detailed dental traits from vertebrates
J. N. Wibisana, R. Sallan, … L. Sallan·bioRxiv·Nov 21, 2023