Correlation of two different devices for the evaluation of primary implant stability depending on dental implant length and bone density: an in vitro study
Abstract
Non-invasive objective implant stability measurements are needed to determine the appropriate timing of prosthetic fitting after implant placement. We compared the early implant stability results obtained using resonance frequency analysis (RFA) and damping capacity analysis (DCA) depending on the implant length and bone density. Total 60, 4.0 mm diameter implants of various lengths (7.3 mm, 10 mm, and 13 mm) were used. Group I and Group II had implants placed in an artificial bone model with a uniform density of 15 PCF (0.24 g/cm3) and 30 PCF (0.48 g/cm3), respectively. RFA was performed using an Osstell(R) Beacon+; DCA was performed using Anycheck(R). Measurements were repeated five times for each implant. Statistical significance was set at P <0.05. In Group I, bone density and primary implant stability were positively correlated, while implant length and primary implant stability were positively correlated. In Group II, the implant stability quotient (ISQ) and implant stability test (IST) values in did not change significantly above a certain length. Primary implant stability was positively correlated with bone density and improved with increasing implant length at low bone densities. Compared with the Osstell(R) Beacon+, the simplicity of Anycheck(R) was easy to use and accessible.
02 /Specification & Classification
Click any tag to browse matching papers- Approach
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Acoustic Prediction and Biomechanical Validation of Primary Stability in Uncemented Short-Stem Hip Prostheses: An In Vitro Study
Alexander Jahnke, Simon Schreynemackers, … Bernd Alexander Ishaque·Journal of Orthopaedic Research®Peer-reviewed·Dec 13, 2025
- Biomechanical Evaluation of Temporomandibular Joint Implants and Periprosthetic Bone under Unilateral and Bilateral Clenching
Rajdeep Ghosh, Girish Chandra, … Kaushik Mukherjee·Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in MedicinePeer-reviewed·May 27, 2025
- Trueness of digital intraoral impression in reproducing multiple implant position
R. J.-Y. Kim, G. I. Benic, J.-M. Park·bioRxiv·Aug 22, 2019
- Core bone diameter in an organic implant-less technique affecting the biomechanical properties of the anterior cruciate ligament fixation; an in-vitro study
M. Mohseni, A. Nourani, … M. Chizari·bioRxiv·Jul 12, 2021
- Validation of a smartphone embedded inertial measurement unit for measuring postural stability in older adults
F. De Groote, S. Vandevyvere, … J.-J. Orban de Xivry·bioRxiv·May 29, 2020
- Evaluation Of Effect Of Bacteria On The Surface Of The Dental Implant
A. S. George, S. S, … T. N·bioRxiv·Jan 5, 2023