Acoustic Transparency Enabling Functional Ultrasound Imaging Through Mouse and Human Skulls
Zikai Wang, Zhengchang KouiD, Junhang ZhangiD, Jingcheng LiiD, Chen GongiD, Chutian Lin, Jincheng WuiD, Michael L. OelzeiD, Qifa ZhouiD, Yaoheng YangiD
Abstract
Ultrasound is one of the most widely used non-invasive imaging modalities for internal organs, yet its application to the brain has been hindered for decades by the skulls acoustic barrier. This same barrier has kept functional ultrasound imaging (fUSI)--an emerging technology capable of capturing brain-wide neural activity in real time at sub-100 m resolution--from reaching its transformative potential in neuroscience and clinical medicine. Here, we present an acoustic transparency strategy by modulating skull acoustic properties, to render the brain visible to ultrasound. A brief topical application of an FDA-approved chelating agent matches the skulls acoustic impedance and sound speed to those of soft tissue, enabling nearly complete ultrasound transmission (94.0 {+/-} 4.4%) with minimal energy loss or distortion. Leveraging this discovery, we developed an acoustic-transparent fUSI platform that maps brain activation across the full brain depth at [~]20 m spatial resolution without skull removal. This method enables safe, longitudinal brain imaging, is fully reversible, and is demonstrated to be applicable to the ex vivo human skull. This conceptually distinct paradigm--controlling acoustic wave propagation via acoustic property modulation--offers a practical and generalizable solution to one of the most persistent obstacles in transcranial ultrasound imaging, opening the door to broader clinical and research application of ultrasound neuroimaging.
02 /Specification & Classification
Click any tag to browse matching papers- Citations
- 3
- Organism
- Assay
- Research Focus
- Approach
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- A window to the brain: ultrasound imaging of human neural activity through a permanent acoustic window
Claire Rabut, Sumner L. Norman, … Mikhail G. Shapiro·Science Translational MedicinePeer-reviewed·May 29, 2024
- Ultrasound Transparent Neural Interfaces for Multimodal Interaction
Raphael Panskus, Andrada Iulia Velea, … Vasiliki Giagka·npj Flexible ElectronicsPeer-reviewed·Jan 7, 2026
- Acoustic properties across the human skull
T. Riis, T. Webb, J. Kubanek·bioRxiv·Apr 22, 2021
- Overcoming the skull barrier for noninvasive transcranial functional ultrasound imaging in marmosets
Hamidreza Ramezanpour, Amirhossein Asadian, … Liya Ma·bioRxiv·Mar 6, 2026
- Adaptive transcranial ultrasound Doppler imaging of the brain
Rick Waasdorp, Eleonora Muñoz-Ibarra, … David Maresca·IEEE Transactions on UltrasonicsPeer-reviewed·Mar 18, 2026
- Full-waveform inversion imaging of the human brain
Lluís Guasch, Òscar Calderón Agudo, … Michael R. Warner·npj Digital MedicinePeer-reviewed·Mar 6, 2020