Application of Spectral Domain Optical Coherence Tomography To Guide Cochlear Implant Electrode Array

Pawina JiramongkolchaiiD, Marcello Magri AmaraliD, Ratul Paul, Abby Matt, Meitong NieiD, Senyue HaoiD, Aaron J AdkinsiD, Hongwu LiangiD, Timothy A. HoldeniD, Craig A. BuchmaniD, Chao ZhouiD

PDFDOI

Abstract

HypothesisA custom spectral domain optical coherence tomography (SD-OCT) platform can be used for real-time guidance of a cochlear implant electrode array (EA). BackgroundWith current cochlear implant surgical techniques, placement of the EA is a blind maneuver in which the surgeon relies on tactile feedback as the EA advances through the cochlear lumen. Cochlear implant trauma is a leading factor for poor speech performance outcomes and loss of residual hearing following surgery. Optical coherence tomography (OCT) is a non-invasive imaging modality that provides real-time visualization of tissue microstructure at higher spatial resolutions compared to clinical CT and MRI. Already adopted as standard of care in ophthalmology, OCT has the potential to assist the surgeon in real-time visualization of the EA trajectory. Unlike commercial systems, our custom OCT system allows tailored wavelength, scanning geometry, and real-time processing, which are critical factors for navigating the compact anatomy of the facial recess to image the cochlea. MethodsA custom-built SD-OCT system was used to image cochlear microanatomy in mice and human cadaveric temporal bones. The OCT system was then used to guide a mock EA in human cadaveric temporal bones in real-time using individual B scans that were reviewed sequentially as the EA was being advanced through the round window. ResultsUsing our OCT system, high-resolution (< 5.0 m) images of cochlear microanatomy were obtained in both mice and cadaveric human temporal bones with an image sensitivity of [~]104 dB. Following cochleostomy in cadaveric temporal bones, real-time sequential OCT B-scans were used to reliably guide placement of the EA through the scala tympani. ConclusionOur custom-built SD-OCT platform can generate high-resolution real-time visualization and orientation of mammalian cochlear microanatomy that can be used to assist with real time guidance of a CI EA. This technology has the potential to serve as a real-time surgical image guidance tool to minimize EA trauma and further our understanding of human cochlear pathophysiology.

Citations
1

Reproducibility ledger

Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.

Sign in to file a reproducibility report.

Similarity Map
SelectedSame fieldCross-field
THISAlbrecht '26Secchia '25Jablonski '25Khurana '26Koert '26Borjigin '26
68% semantic matchNeuroscience
Niels Albrecht, Elisabeth Koert, Anna Vavakou+10 more·bioRxiv·Sep 1, 2026