Attenuation of Motion Artifacts in fMRI using Discrete Reconstruction of Irregular fMRI Trajectories (DRIFT)
David B. ParkeriD, Pascal SpincemailleiD, Qolamreza Ray RazlighiiD
Abstract
Numerous studies reported motion as the most detrimental source of noise and artifacts in functional magnetic resonance imaging (fMRI). Different approaches have been proposed and used to attenuate the effect of motion on fMRI data, including both prospective and retrospective (post-processing) techniques. However, each type of motion (e.g. translation versus rotation or in-plane versus out-of-plane) has a distinct effect on the MR signal, which is not fully understood nor appropriately modeled in the field. In addition, effects of the same motion can be substantially different depending on when it occurs during the pulse sequence (e.g. RF excitation, gradient encoding, or k-space read-out). Thus, each distinct kind of motion and the time of its occurrence may require a unique approach to be optimally corrected. Therefore, we start with an investigation of the effects of different motions on the MR signal based on the Bloch equation. We then simulate their unique effects with a comprehensive fMRI simulator. Our results indicate that current motion correction methods fail to completely address the motion problem. Retrospective techniques such as spatial realignment can correct for between-volume misalignment, but fail to address within volume contamination and spin-history artifacts. Because of the steady state nature of the fMRI acquisition, spin-history artifacts arising from over/under excitation during slice-selection causes the motion artifacts to contaminate MR signal even after cessation of motion, which makes it challenging to be corrected retrospectively. Prospective motion correction has been proposed to prevent spin-history artifacts, but fails to address motion artifacts during k-space readout. In this article, we propose a novel method to remove these artifacts: Discrete reconstruction of irregular fMRI trajectory (DRIFT). Our method calculates the exact displacement of k-space recording due to motion at each dwell time and retrospectively corrects each slice of the fMRI volume using an inverse nonuniform Fourier transform. We evaluate our proposed methods using simulated data as well as fMRI data collected from a rotating phantom inside a 3T Siemens Prisma scanner. We conclude that a hybrid approach with both prospective and retrospective components are essentially required for optimal removal of motion artifacts from the fMRI data.
- Citations
- 2
- FWCI
- 0.26
- Field percentile
- 48th percentile
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Inter-slice motion correction using spatiotemporal interpolation for functional magnetic resonance imaging of the moving fetus
Wonsang You, Catherine Limperopoulos·bioRxiv·Oct 17, 2017
- Correction of respiratory artifacts in MRI head motion estimates
Damien A. Fair, Óscar Miranda-Domínguez, … Nico U.F. Dosenbach·NeuroImagePeer-reviewed·Nov 25, 2019
- Feasibility of spiral fMRI based on an LTI gradient model
Nadine N. Graedel, Lars Kasper, … S. Johanna Vannesjo·NeuroImagePeer-reviewed·Oct 28, 2021
- Modular preprocessing pipelines can reintroduce artifacts into fMRI data
Martin A. Lindquist, Stephan Geuter, … Brian S. Caffo·Human Brain MappingPeer-reviewed·Jan 21, 2019
- Removal of high frequency contamination from motion estimates in single-band fMRI saves data without biasing functional connectivity
Caterina Gratton, Ally Dworetsky, … Meghan C. Campbell·NeuroImagePeer-reviewed·Apr 20, 2020
- Evaluation and Optimization of Motion Correction in Spinal Cord fMRI Preprocessing
Hamed Dehghani, Kenneth Arnold Weber II, … Ali Khatibi·bioRxiv·May 20, 2020