Validity of the Lumen(R) hand-held metabolic device to measure fuel utilization in healthy young adults
K. A. Lorenz, S. Yeshurun, R. Aziz, J. Ortiz-Delatorre, J. R. Bagley, M. Mor, M. Kern
Abstract
BackgroundMetabolic carts measure the carbon dioxide produced and oxygen consumed from the breath in order to assess metabolic fuel usage (carbohydrates vs. fats). However, these systems are expensive, time-consuming, and only available in the clinic. A small hand-held device capable of measuring metabolic fuel via CO2 from exhaled air has been developed ObjectiveTo evaluate the validity of a novel hand-held device (Lumen(R)) for measuring metabolic fuel utilization in healthy young adults MethodsMetabolic fuel usage was assessed in healthy participants (n = 33; age: 23.1 {+/-} 3.9 y) via respiratory exchange ratio (RER) values from the "gold-standard" metabolic cart as well as %CO2 from the Lumen device. Measurements were performed at rest in two conditions, fasting, and after consuming 150 grams of glucose in order to determine changes in metabolic fuel. Reduced major axis regression was performed as well as Bland-Altman plots and linear regressions to test for agreement between RER and Lumen %CO2. ResultsBoth RER and Lumen %CO2 significantly increased after glucose intake compared with fasting conditions (P < .0001). Regression analyses and Bland-Altman plots revealed an agreement between the two measurements with a systematic bias resulting from the nature of the different units. ConclusionsThis study shows the validity of Lumen(R) to estimate metabolic fuel utilization in a comparable manner with the "gold-standard" metabolic cart, providing the ability for real-time metabolic information for users under any circumstances.
- organism: human
- assay: metabolomics
- research focus: drug response / drug discovery
- approach: statistical model
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Low Cost Metabolic Fuel Sensor for On-Demand Personal Health and Fitness Tracking
L. M. Candell, R. W. Hoyt, … G. A. Shaw·bioRxiv·Feb 20, 2020
- Validity and reliability of a new whole room indirect calorimeter to assess metabolic response to small-calorie loads
M. E. Baugh, M. L. Ahrens, … A. G. DiFeliceantonio·bioRxiv·Sep 21, 2023
- Recurrence quantification analysis of heart rate variability is a COVID-safe alternative to gas analysis in the detection of metabolic thresholds
G. Zimatore, L. Falcioni, … C. Baldari·bioRxiv·Mar 22, 2021
- Optimization and Validation of Whole-Room Indirect Calorimetry for Improved Accuracy and Temporal Resolution
Christopher Hayden, Luke R. Arieta, … Michael A. Busa·bioRxiv·Apr 14, 2026
- Methodologic Issues in Doubly Labeled Water Measurements of Energy Expenditure During Very Low-Carbohydrate Diets
K. D. Hall, J. Guo, … E. Ravussin·bioRxiv·Oct 23, 2018
- Glucose Uptake as an Alternative to Stop-Flow Respirometry for Measuring Metabolic Rate in Danio rerio Larvae
B. L. Evans, H. A. Shiels, … P. E. Clayton·bioRxiv·Apr 30, 2021