A comprehensive evaluation of multiband-accelerated sequences and their effects on statistical outcome measures in fMRI.
Lysia Demetriou, Oliwia S Kowalczyk, Gabriella Tyson, Thomas Bello, Rexford D Newbould, Matthew B Wall
Abstract
Accelerated functional Magnetic Resonance Imaging (fMRI) with multiband sequences is now relatively widespread. These sequences can be used to dramatically reduce the repetition time (TR) and produce a time-series sampled at a higher temporal resolution. We tested the effects of higher temporal resolutions for fMRI on statistical outcome measures in a comprehensive manner on two different MRI scanner platforms. Experiment 1 tested a range of acceleration factors (1-6) against a standard EPI sequence on a single composite task that maps a number of basic sensory, motor, and cognitive networks. Experiment 2 compared the standard sequence with acceleration factors of 2 and 3 on both resting-state and two task paradigms (an N-back task, and faces/places task), with a number of different analysis approaches. Results from experiment 1 showed modest but relatively inconsistent effects of the higher sampling rate on statistical outcome measures. Experiment 2 showed strong benefits of the multiband sequences on results derived from resting-state data, but more varied effects on results from the task paradigms. Notably, the multiband sequences were superior when Multi-Voxel Pattern Analysis was used to interrogate the faces/places data, but showed less benefit in conventional General Linear Model analyses of the same data. In general, ROI-derived measures of statistical effects benefitted relatively little from higher sampling resolution, with decrements even seen in one task (N-back). Across both experiments, results from the two scanner platforms were broadly comparable. The statistical benefits of high temporal resolution fMRI with multiband sequences may therefore depend on a number of factors, including the nature of the investigation (resting-state vs. task-based), the experimental design, the particular statistical outcome measure, and the type of analysis used. Higher sampling rates in fMRI are not a panacea, and it is recommended that researchers use multiband acquisition sequences conservatively.
02 /Specification & Classification
Click any tag to browse matching papersReproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Does higher sampling rate (Multiband + SENSE) benefit the detection of task correlated BOLD for cognitive neuroscience applications at 3T?
Ritu Bhandari, Evgeniya P. Kirilina, … Valeria Gazzola·bioRxiv·Sep 9, 2019
- Statistical Power or More Precise Insights into Neuro-Temporal Dynamics? Assessing the Benefits of Rapid Temporal Sampling in fMRI
Logan Thorne Dowdle, Geoffrey M. Ghose, … Luca Vizioli·Progress in NeurobiologyPeer-reviewed·Sep 4, 2021
- Multiband acceleration can provide moderate improvements in single-subject voxel-wise statistics in block-design task-based fMRI
Ritu Bhandari, Valeria Gazzola, Christian Keysers·bioRxiv·May 26, 2020
- Comparing the effect of multi gradient echo and multi band fMRI during a semantic task
Ajay D. Halai, Richard N. Henson, … Marta Correia·Imaging NeurosciencePeer-reviewed·Jan 1, 2025
- The effects of multi-echo fMRI combination and rapid T2*-mapping on offline and real-time BOLD sensitivity
Stephan Heunis, Marcel M. Breeuwer, … Albert Pierre Aldenkamp·NeuroImagePeer-reviewed·Jun 8, 2021
- Improving the temporal accuracy of functional Magnetic Resonance Imaging
Niels Janssen, Juan A. Hernández-Cabrera, Laura Ezama Foronda·bioRxiv·Sep 5, 2016