bioRxivNeurosciencePreprint
Recovery Dynamics of the High Frequency Alternating Current Nerve Block
Adrien RapeauxiD, Emma BruntoniD, Kianoush NazarpouriD, Timothy G. ConstandinouiD
Abstract
ObjectiveHigh-Frequency alternating current (HFAC) nerve block has great potential for neuromodulation-based therapies. However nerve function recovery dynamics after a block is highly understudied. This study aims to characterise the recovery dynamics of neural function after an HFAC block.\n\nApproachExperiments were carried out in-vivo to determine blocking efficacy as a function of blocking signal amplitude and frequency, and recovery times as well as recovery completeness was measured within a 0.7 s time scale from the end of block. The sciatic nerve was stimulated at 100 Hz during recovery to reduce error to within {+/-}10 ms for measurements of recovery dynamics. The electromyogram (EMG) signals were measured from gastrocnemius medialis and tibialis anterior during trials as an indicator for nerve function.\n\nMain ResultsThe HFAC block was most reliable around 20 kHz, with block thresholds approximately 5 or 6 mA depending on the animal and muscle. Recovery times ranged from 20 to 430 milliseconds and final values spanned relative outputs from approximately 1 to 0.2. Higher blocking signal frequencies and amplitudes increased recovery time and decreased recovery completeness.\n\nSignificanceThese results confirm that recovery dynamics from block depend on blocking signal frequency and amplitude, which is of particular importance for neuromodulation therapies and for comparing results across studies using different blocking signal parameters.
- Citations
- 2
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Precision cooling assists electrical stimulation: Breaking selectivity barriers in nerve control
Shu Yang, Xiao Yu, … Zhanhong Jeff Du·bioRxiv·Sep 16, 2026
- A partially averaged system to model neuron responses to interferential current stimulation
Eduardo Cerpa, Matías Courdurier, … Esteban Paduro·Journal of Mathematical BiologyPeer-reviewed·Dec 5, 2022
- Fundamental limitations of kilohertz-frequency carriers in afferent fiber recruitment with transcutaneous spinal cord stimulation
Rodolfo Keesey, Ursula S. Hofstoetter, … Ismael Seáñez·bioRxiv·Jul 26, 2024
- A Block-Capable and Module-Extendable 4-Channel Stimulator for Acute Neurophysiology
A. Rapeaux, T. Constandinou·bioRxiv·Jan 28, 2020
- Towards a more accurate quasi-static approximation of the electric potential for neurostimulation with kilohertz-frequency sources
Thomas Caussade, Esteban Paduro, … Leonel E. Medina·Journal of Neural EngineeringPeer-reviewed·Dec 1, 2023
- Dynamical Diversity in Conductance-Based Neuron Response to kilohertz Electrical Stimulation
J. G. Polli, Florian Kölbl, … M. G. E. da Luz·Physical review. EPeer-reviewed·Mar 18, 2026