A Personalized Switched Systems Approach for the Optimal Control of Ventricular Assist Devices based on Atrioventricular Plane Displacement
C. Zeile, T. Rauwolf, A. Schmeisser, J. K. Mizerski, R. C. Braun-Dullaeus, S. Sager
Abstract
ObjectiveA promising treatment for congestive heart failure is the implementation of a left ventricular assist device (LVAD) that works as a mechanical pump. Modern LVADs work with adjustable constant rotor speed and provide therefore continuous blood flow; however, recently undertaken efforts try to mimic pulsatile blood flow by oscillating the pump speed. This work proposes an algorithmic framework to construct and evaluate optimal pump speed policies. MethodsWe use a model that captures the atrioventricular plane displacement, which is a physiological indicator for heart failure. We employ mathematical optimization to adapt this model to patient specific data and to find optimal pump speed policies with respect to ventricular unloading and aortic valve opening. To this end, we reformulate the cardiovascular dynamics into a switched system and thereby reduce nonlinearities. We consider system switches that stem from varying the constant pump speed and that are state dependent such as valve opening or closing. ResultsAs a proof of concept study, we personalize the model to a selected patient with respect to ventricular pressure. The model fitting results in a root-mean-square deviation of about 6 mmHg. Optimized constant and piecewise constant rotor speed profiles improve the default initialized solution by 31% and 68% respectively. ConclusionThese in silico findings demon-strate the potential of personalized hemodynamical optimization for the LVAD therapy. SignificanceLVADs and their optimal configuration are active research fields. Mathematical optimization enhances our understanding of how LVADs should provide pulsatility.
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- A New Mathematical Model for LVAD-Supported Ventricles: Direct Parameterization from Ramp-Test Clinical Data and Verification via Hybrid Modeling
E. Umo Abraham, A. Welch Brett, … Ethan O. Kung·bioRxiv·Mar 23, 2026
- Effect of percutaneous ventricular assisted device on post-cardiac arrest myocardial dysfunction in swine model with prolonged cardiac arrest
T. Nakashima, M. H. Tiba, … R. W. Neumar·ResuscitationPeer-reviewed·Dec 1, 2023
- Design and execution of a Verification, Validation, and Uncertainty Quantification plan for a numerical model of left ventricular flow after LVAD implantation
A. Santiago, C. Butakoff, … M. Vazquez·bioRxiv·Nov 11, 2021
- Data Driven Control of Vagus Nerve Stimulation for the Cardiac System
- Design and Validation of an Open Loop Controlled Positive-Displacement Pump for Vascular Flow Modelling
T. O'Brien·bioRxiv·Feb 17, 2023
- The Role of TandemHeartTM combined with ProtekDuoTM as Right Ventricular Support Device: A Simulation Approach
Beatrice De Lazzari, Roberto Badagliacca, … Claudio De Lazzari·Computer Methods and Programs in BiomedicinePeer-reviewed·Oct 28, 2024