Design and execution of a Verification, Validation, and Uncertainty Quantification plan for a numerical model of left ventricular flow after LVAD implantation
A. Santiago, C. Butakoff, B. Eguzkitza, R. A. Gray, K. May-Newman, P. Pathmanathan, V. Vu, M. Vazquez
Abstract
Backgroundleft ventricular assist devices (LVADs) are implantable pumps that act as a life support therapy for patients with severe heart failure. Despite improving the survival rate, LVAD therapy can carry major complications. Particularly, the flow distortion introduced by the LVAD in the left ventricle (LV) may induce thrombus formation. While previous works have used numerical models to study the impact of multiple variables in the intra-LV stagnation regions, a comprehensive validation analysis has never been executed. The main goal of this work is to present a model of the LV-LVAD system and to design and follow a verification, validation and uncertainty quantification (VVUQ) plan based on the ASME V&V40 and V&V20 standards to ensure credible predictions. MethodsThe experiment used to validate the simulation is the SDSU cardiac simulator, a bench mock-up of the cardiovascular system that allows mimicking multiple operation conditions for the heart-LVAD system. The numerical model is based on Alya, the BSCs in-house platform for numerical modelling. Alya solves the Navier-Stokes equation with an Arbitrarian Lagrangian-Eulerian (ALE) formulation in a deformable ventricle and includes pressure-driven valves, a 0D Windkessel model for the arterial output and a LVAD boundary condition modeled through a dynamic pressure-flow performance curve. The designed VVUQ plan involves: (a) a risk analysis and the associated credibility goals; (b) a verification stage to ensure correctness in the numerical solution procedure; (c) a sensitivity analysis to quantify the impact of the inputs on the four quantities of interest (QoIs) (average aortic root flow [Formula], maximum aortic root flow [Formula], average LVAD flow [Formula], and maximum LVAD flow [Formula]); (d) an uncertainty quantification using six validation experiments that include extreme operating conditions. ResultsNumerical code verification tests ensured correctness of the solution procedure and numerical calculation verification showed small numerical errors. The total Sobol indices obtained during the sensitivity analysis demonstrated that the ejection fraction, the heart rate, and the pump performance curve coefficients are the most impactful inputs for the analysed QoIs. The Minkowski norm is used as validation metric for the uncertainty quantification. It shows that the midpoint cases have more accurate results when compared to the extreme cases. The total computational cost of the simulations was above 100 [core-years] executed in around three weeks time span in Marenostrum IV supercomputer. ConclusionsThis work details a novel numerical model for the LV-LVAD system, that is supported by the design and execution of a VVUQ plan created following recognised international standards. We present a methodology demonstrating that stringent VVUQ according to ASME standards is feasible but computationally expensive.
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Blood flow modeling under LVAD physiology. From global circulation to local hemodynamics
P. J. Blanco, J. Grinstein, … H. Garcia-Garcia·bioRxiv·Jan 28, 2022
- A Personalized Switched Systems Approach for the Optimal Control of Ventricular Assist Devices based on Atrioventricular Plane Displacement
C. Zeile, T. Rauwolf, … S. Sager·bioRxiv·May 27, 2020
- A coupled atrioventricular-aortic setup for in-vitro hemodynamic study of the systemic circulation: Design, Fabrication, and Physiological relevancy
R. Alavi, A. Aghilinejad, … N. M. Pahlevan·bioRxiv·Apr 18, 2022
- Modelling and simulation of the interactions between the cardiovascular system, VA ECMO and IABP: when is it an appropriate combination?
Beatrice De Lazzari, Massimo Capoccia, … Claudio De Lazzari·bioRxiv·Feb 3, 2025
- A New Mathematical Model for LVAD-Supported Ventricles: Direct Parameterization from Ramp-Test Clinical Data and Verification via Hybrid Modeling
E. Umo Abraham, A. Welch Brett, … Ethan O. Kung·bioRxiv·Mar 23, 2026
- Calibration and validation strategy for electromechanical cardiac digital twins
Zhinuo Jenny Wang, Maxx Holmes, … Blanca Rodríguez·bioRxiv·Mar 6, 2025