Accuracy and Reliability of a Suite of Digital Measures of Walking Generated Using a Wrist-worn Sensor: Performance Characterization Study
Nathan KowahliD, Sooyoon ShiniD, Poulami BarmaniD, Erin RainaldiiD, Sara F. PophamiD, Ritu KapuriD
Abstract
Background/ObjectivesMobility is a meaningful aspect of an individuals health whose quantification can provide clinical insights. Wearable sensor technology can quantify walking behaviors (a key aspect of mobility) through continuous passive monitoring. Our objective was to characterize the accuracy and reliability performance of a suite of digital measures of walking behaviors, as critical aspects in the practical implementation of digital measures into clinical studies. MethodsWe collected data from a wrist-worn device (the Verily Study Watch) worn for multiple days by a cohort of volunteer participants without history of gait/walking impairment in a real world setting. Based on step measurements computed in 10-second epochs from sensor data, we generated individual daily aggregates (participant-days) to derive a suite of measures of walking: step count, walking bout duration, number of total walking bouts, number of long walking bouts, number of short walking bouts, peak 30-minute walking cadence, peak 30-minute walking pace. To characterize accuracy of the measures, we examined agreement with truth labels generated by a concurrent, ankle-worn, reference device (Modus StepWatch 4) with known low error, calculating the following metrics: Intraclass Correlation Coefficient (ICC), Pearson R, Mean Error (ME), Mean Absolute Error (MAE). To characterize the reliability, we developed a novel approach to identify the time to reach a reliable readout (time-to-reliability) for each measure. This was accomplished by computing mean values over aggregation scopes ranging from 1-30 days, and analyzing test-retest reliability based on ICCs between adjacent (non-overlapping) time windows for each measure. ResultsIn the accuracy characterization, we collected data for a total of 162 participant-days from a testing cohort (N=35 participants; median observation time, 5 days). Agreement with the reference device-based readouts in the testing subcohort (n=35) for the eight measurements under evaluation, as reflected by ICCs, ranged between 0.7-0.9; Pearson R values were all greater than 0.75. For the time-to-reliability characterization, we collected data for a total of 15,120 participant days (overall cohort N=234; median observation time, 119 days). Here, all digital measures achieved an ICC between adjacent readouts > 0.75 by 16 days of wear time. ConclusionsWe characterized accuracy and reliability of a suite of digital measures that provides comprehensive information about walking behaviors in real-world settings. These results, which report the level of agreement with high-accuracy reference labels and the time duration required to establish reliable measure readouts, can guide practical implementation of these measures into clinical studies. Well-characterized tools to quantify walking behaviors in research contexts can provide valuable clinical information about general population cohorts and patients with specific conditions.
02 /Specification & Classification
Click any tag to browse matching papers- Citations
- 0
- Assay
- Research Focus
- Approach
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Ambulatory physiological measures obtained under naturalistic urban mobility conditions have acceptable reliability
Dilber Korkmaz, Kilian Knauth, … Jan Peters·Scientific ReportsPeer-reviewed·Aug 7, 2025
- Novel algorithms deriving clinical performance measures from smartphone sensor data collected under a walking test
M. A. Little, S. Volotinen, … B. Byrom·bioRxiv·Oct 21, 2021
- Accuracy in wrist-worn, sensor-based measurements of heart rate and energy expenditure in a diverse cohort
Anna Shcherbina, C. Mikael Mattsson, … Euan Angus Ashley·Journal of Personalized MedicinePeer-reviewed·May 24, 2017
- On the reliability of wearable technology: A tutorial on measuring heart rate and heart rate variability in the wild
Veronica Dudarev, Oswald Barral, … James T. Enns·SensorsPeer-reviewed·Jun 24, 2023
- Behaviour-based movement cut-off points in 3-year old children comparing wrist- with hip-worn actigraphs MW8 and GT3X
Daniel Jansson, Rikard Westlander, … Katharina Wulff·PLoS ONEPeer-reviewed·Mar 26, 2025
- Validation and optimisation of wearable accelerometer data pre-processing for digital measure implementation and development
Joss Langford, Jia Ying Chua, … Melvyn M. Hillsdon·bioRxiv·Apr 6, 2026