Preclinical Evaluation of a Zotarolimus-Eluting Stent for Intracranial Application: A 30-Day GLP Canine Study
Ibrahim Ahmad BhattiiD, Fareed JumahiD, Syed A GillaniiD, Ravi S. NunnaiD, Inamullah Khan, Austin Rolfes, Bharathi JagadeesaniD, Ameer E HassaniD, Adnan Iqbal QureshiiD, Farhan SiddiqiD
Abstract
BackgroundZotarolimus-Eluting Stents (ZES) are increasingly used off-label for intracranial atherosclerotic disease (ICAD). However, foundational, Good Laboratory Practice (GLP) compliant preclinical safety data specific to the cerebral vasculature is lacking. Therefore, we conducted the first systematic evaluation of the technical feasibility, anatomic features, and 30-day clinical safety profile of the Onyx Frontier ZES (Medtronic, Minneapolis, MN) in a GLP-compliant canine intracranial model. MethodsIn a prospective GLP study, a 2.0 x 8.0 mm Onyx Frontier ZES was implanted in the basilar artery of twelve healthy mongrel canines pre-treated with dual antiplatelet therapy. Vertebral artery tortuosity was assessed via angiography and graded as Type I (non-tortuous/straight), Type II (moderately tortuous/S-shaped), or Type III (severely tortuous/coiled). The primary endpoints were procedural success and 30-day safety, assessed via angiography, serial neurological examinations, and comprehensive clinical monitoring. ResultsTechnical success was achieved in 100% of animals (12/12). The cohort exhibited challenging anatomy, with a mean basilar artery diameter of 1.39 {+/-} 0.10 mm. Tortuosity analysis revealed that 33.3% of animals were Type I (mild), 58.3% were Type II (moderate), and 8.3% were Type III (severe). Post-procedure, mild vasospasm was observed in all cases, with only one canine requiring intra-arterial verapamil. A single case (8.3%) of self-limiting contrast extravasation occurred without clinical sequelae. At 30 days, there was no mortality, and no animals suffered a neurological adverse event. Minor clinical events were unrelated to the device, and comprehensive clinical pathology panels revealed no evidence of systemic toxicity or a significant inflammatory response. ConclusionThis first-of-its-kind preclinical study demonstrates that implantation of the Onyx Frontier ZES is technically feasible with a 100% success rate in a sub-2mm vessel model with significant tortuosity and has an excellent 30-day safety profile with no clinical evidence of neurotoxicity. These findings provide foundational preclinical data supporting further investigation of this technology for treating ICAD in human subjects.
02 /Specification & Classification
Click any tag to browse matching papers- Citations
- 0
- Approach
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- A protocol for the reduction of applied torque on parent vessel duringelastase-induced aneurysm formation using rabbit animal models
S. A. Pineda-Castillo, K. A. Laurence, … C.-H. Lee·bioRxiv·Oct 8, 2021
- Atlas drug-eluting coronary stents inhibits neointimal hyperplasia in sheep modeling
R. Dinc, H. Yerebakan·bioRxiv·Apr 19, 2023
- Cell-cultured PDMS vascular model to allow placement of implant devices
Taku Okuno, Kazuyo Ito, … Daisuke Yoshino·BMC MethodsPeer-reviewed·May 11, 2026
- Development of a novel murine model of in-stent neoatherosclerosis
Jiandi Liu, Demeke Geremew, … Christina A. Bursill·Journal of the American Heart AssociationPeer-reviewed·Nov 12, 2025
- Cerebral vascular tortuosity and aneurysm formation and rupture: a novel vessel tortuosity scale
Maxon V. Knott, Naser Hamad, … Robert M Starke·Experimental NeurologyPeer-reviewed·Oct 21, 2025
- Vasoconstriction in isolated goat aorta does not increase mean aortic pressure
N. Gangadharan, A. V, … S. Subramani·bioRxiv·Jan 30, 2021