A new approach to automated CBMN scoring following high doses

M. Repin, D. J. Brenner, G. Garty

Abstract

In recent years we have automated the CBMN assay using microvolumes of blood, processed in multiwell plates. We have seen that at doses above 6 Gy the detected yield of micronuclei actually declines with dose, likely because of mitotic delay, preventing cells from forming micronuclei and also, when using one color imaging, resulting in many false binucleated cells, consisting of two randomly-adjacent nuclei. By using the inverse mitotic index (the ratio of mononuclear to binuclear cells) to adjust the micronucleus yield we were able to obtain a monotonic increasing dose response curve at doses of up to at least 10 Gy from the same samples which generated dose-response curve with a peak near 6 Gy, when scored using the traditional micronucleus yield.