bioRxivBioengineering
Fluid-structure interaction analysis of eccentricity and leaflet rigidity on thrombosis biomarkers in bioprosthetic aortic valve replacements
Abstract
This work introduces the first 2-way fluid-structure interaction (FSI) computational model to study the effect of aortic annulus eccentricity on the performance and thrombogenic risk of cardiac bioprostheses. The model predicts that increasing eccentricities yield lower geometric orifice areas (GOAs) and higher normalized transvalvular pressure gradients (TPGs) for healthy cardiac outputs during systole, agreeing with in vitro experiments. Regions with peak values of residence time and shear rate are observed to grow with eccentricity in the sinus of Valsalva, indicating an elevated risk of thrombus formation for eccentric configurations. In addition, the computational model is used to analyze the effect of varying leaflet rigidity on both performance, thrombogenic and calcification risks with applications to tissue-engineered prostheses, observing an increase in systolic and diastolic TPGs, and decrease in systolic GOA, which translates to decreased valve performance for more rigid leaflets. An increased thrombogenic risk is detected for the most rigid valves. Peak solid stresses are also analyzed, and observed to increase with rigidity, elevating risk of valve calcification and structural failure. The immersed FSI method was implemented in a high-performance computing multi-physics simulation software, and validated against a well known FSI benchmark. The aortic valve bioprosthesis model is qualitatively contrasted against experimental data, showing good agreement in closed and open states. To the authors knowledge this is the first computational FSI model to study the effect of eccentricity or leaflet rigidity on thrombogenic biomarkers, providing a novel tool to aid device manufacturers and clinical practitioners.
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- An accelerated thrombosis model for computational fluid dynamics simulations in rotary blood pumps
C. Blum, S. Gross-Hardt, … M. Neidlin·bioRxiv·Aug 30, 2021
- Capturing Multi-Scale Dynamics of Aortic Valve Calcification With a Coupled Fluid Structure and Systems Biology Model
Michael Quan, Tianyou Xie, … Haoxiang Luo·bioRxiv·Mar 19, 2026
- Longitudinal Investigation of Aortic Dissection in Mice with Computational Fluid Dynamics
K. Baeumler, E. H. Phillips, … C. J. Goergen·bioRxiv·Apr 24, 2023
- Effects of cardiac function alterations on the risk of postoperative thrombotic complications in patients receiving endovascular aortic repair
- Aortic annuloplasty FSI digital twin of 3D-printed phantoms with 4D-flow MRI comparison
Luca Bontempi, Marta Zattoni, … Monika Colombo·Computers in Biology and MedicinePeer-reviewed·May 17, 2025
- Predicting post-TEVAR endoleaks: a pre-operative hemodynamic risk factor from patient-specific Fluid-Structure Interaction simulations
Francesca Duca, Silvia Tavarone, … Francesco Migliavacca·Annals of Biomedical EngineeringPeer-reviewed·Oct 27, 2026