bioRxivBioengineering
First-in-human demonstration of floating EMG sensors and stimulators wirelessly powered and operated by volume conduction
L. Becerra-Fajardo, J. Minguillon, M. O. Krob, C. Rodrigues, M. Gonzalez-Sanchez, A. Megia-Garcia, C. Redondo Galan, F. Gutierrez Henares, A. Comerma, A. J. del-Ama, A. Gil-Agudo, F. Grandas, A. Schneider-Ickert, F. Oliveira Barroso, A. Ivorra
Abstract
BackgroundRecently we reported the design and evaluation of floating semi-implantable devices that receive power from and bidirectionally communicate with an external system using coupling by volume conduction. The approach, of which the semi-implantable devices are proof-of-concept prototypes, may overcome some limitations presented by existing neuroprostheses, especially those related to implant size and deployment, as the implants avoid bulky components and can be developed as threadlike devices. Here, it is reported the first-in-human acute demonstration of these devices for electromyography (EMG) sensing and electrical stimulation. MethodsA proof-of-concept device, consisting of implantable thin-film electrodes and a non-implantable miniature electronic circuit connected to them, was deployed in the upper or lower limb of six healthy participants. Two external electrodes were strapped around the limb and were connected to the external system which delivered high frequency current bursts. Within these bursts, 13 commands were modulated to communicate with the implant. ResultsFour devices were deployed in the biceps brachii and the gastrocnemius medialis muscles, and the external system was able to power and communicate with them. Limitations regarding insertion and communication speed are reported. Sensing and stimulation parameters were configured from the external system. In one participant, electrical stimulation and EMG acquisition assays were performed, demonstrating the feasibility of the approach to power and communicate with the floating device. ConclusionsThis is the first-in-human demonstration of EMG sensors and electrical stimulators powered and operated by volume conduction. These proof-of-concept devices can be miniaturized using current microelectronic technologies, enabling fully implantable networked neuroprosthetics.
- organism: human
- assay: electrophysiology
- research focus: drug response / drug discovery
- approach: experimental assay
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Floating EMG Sensors and Stimulators Wirelessly Powered and Operated by Volume Conduction for Networked Neuroprosthetics
L. Becerra-Fajardo, M. O. Krob, … A. Ivorra·bioRxiv·Mar 9, 2022
- Wireless networks of injectable microelectronic stimulators based on rectification of volume conducted high frequency currents
A. Garcia-Moreno, A. Comerma-Montells, … A. Ivorra·bioRxiv·Mar 11, 2022
- Powering electronic implants by high frequency volume conduction: in human validation
J. Minguillon, M. Tudela-Pi, … A. Ivorra·bioRxiv·Mar 15, 2021
- Wireless endovascular nerve stimulation with a millimeter-sized magnetoelectric implant
J. C. Chen, P. Kan, … J. T. Robinson·bioRxiv·Jul 6, 2021
- Wireless Ensembles of Sub-mm Microimplants Communicating as a Network near 1 GHz in a Neural Application
Ji-Hun Lee, Vincent Leung, … Arto V Nurmikko·bioRxiv·Sep 11, 2020
- A Fully Endovascular Neural Interface
John Stanton, Giovanni Talei Franzesi, … Kenneth L. Shepard·bioRxiv·Jun 11, 2026