bioRxivBiophysicsPreprint
An In-Silico Testbed for Fast and Accurate MR Labeling of Orthopaedic Implants
G. M. Noetscher, P. J. Serano, M. Horner, A. Prokop, J. Hanson, K. Fujimoto, J. E. Brown, A. Nazarian, J. Ackerman, S. N. Makaroff
Abstract
AO_SCPLOWBSTRACTC_SCPLOWOne limitation on the ability to monitor health in older adults using Magnetic Resonance (MR) imaging is the presence of implants, where the prevalence of implantable devices (orthopedic, cardiac, neuromodulation) increases in the population, as does the pervasiveness of conditions requiring MRI studies for diagnosis (musculoskeletal diseases, infections, or cancer). The present study describes a novel multiphysics implant modeling testbed using the following approaches with two examples: - an in-silico human model based on the widely available Visible Human Project (VHP) cryo-section dataset; - a finite element method (FEM) modeling software workbench from Ansys (Electronics Desktop/Mechanical) to model MR radio frequency (RF) coils and the temperature rise modeling in heterogeneous media. The in-silico VHP Female model (250 parts with an additional 40 components specifically characterizing embedded implants and resultant surrounding tissues) corresponds to a 60-year-old female with a body mass index (BMI) of 36. The testbed includes the FEM-compatible in-silico human model, an implant embedding procedure, a generic parameterizable MRI RF birdcage two-port coil model, a workflow for computing heat sources on the implant surface and in adjacent tissues, and a thermal FEM solver directly linked to the MR coil simulator to determine implant heating based on an MR imaging study protocol. The primary target is MR labeling of large orthopaedic implants. The testbed has very recently been approved by the US Food and Drug Administration (FDA) as a medical device development tool (MDDT) for 1.5 T orthopaedic implant examinations.
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Application of Surgical Lead Management and Reconfigurable Coil Technology to Reduce RF Heating of DBS Implants during MRI at 3T Under Variant Body Compositions
B. Bhusal, B. Elahi, … L. Golestani Rad·bioRxiv·Jun 25, 2020
- RF heating of deep brain stimulation implants in open-bore vertical MRI systems
L. Golestani Rad, D. Lampman, … J. Kirsch·bioRxiv·May 27, 2019
- RF Heating of Bipolar Epicardial Implants during MRI at 0.55 T and 1.5T: Effect of Device Length and Termination Conditions
Bhumi Bhusal, Pia Sanpitak, … Laleh Golestanirad·bioRxiv·May 29, 2026
- Age and lead configuration matter: A comparative study of RF-induced heating of epicardial and endocardial electronic devices in adult and pediatric anthropomorphic phantoms in 1.5 T MR
- Modifying the trajectory of epicardial leads can substantially reduce MRI-induced RF heating in pediatric patients with a cardiac implantable electronic device
F. Jiang, B. Bhusal, … L. Golestanirad·bioRxiv·Jul 26, 2022
- Reconfigurable MRI coil technology can substantially reduce RF heating at the tips of bilateral deep brain stimulation implants
L. Golestanirad, B. Keil, … L. L. Wald·bioRxiv·Nov 22, 2018