Application of Surgical Lead Management and Reconfigurable Coil Technology to Reduce RF Heating of DBS Implants during MRI at 3T Under Variant Body Compositions
B. Bhusal, B. Elahi, B. Keil, J. Rosenow, E. Kazemivalipour, L. Golestani Rad
Abstract
Patients with active implants such as deep brain stimulation (DBS) devices, have limited access to magnetic resonance imaging (MRI) due to risks of RF heating. With an aging population, the prevalence of neurodegenerative and vascular disease increases; and so does the indication for MRI exams in patients with such implants. In response to this growing need for MRI, many groups have investigated strategies to mitigate the RF heating of the implants. These efforts, however, have relied either on simulations with homogenous body models or simplified phantom experiments (box shaped phantom with single tissue). It is well established, however, that the shape and heterogeneity of human body affects the distribution of MRI electric fields, which by proxy, alters the RF heating of an implant inside the body. In this contribution, we applied numerical simulations and phantom experiments to examine the effectiveness of RF heating mitigation strategies under variant patient body compositions, focusing on two recently proposed techniques: (a) surgical modification of DBS lead trajectories inside the body, and (b) use of a patient-adjustable reconfigurable MRI coil, both aiming to reduce the coupling of implanted leads and MRI electric fields. Our results demonstrated that both techniques perform well under variant body compositions.Competing Interest StatementThe authors have declared no competing interest.View Full Text
02 /Specification & Classification
Click any tag to browse matching papers- Approach
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Reconfigurable MRI coil technology can substantially reduce RF heating at the tips of bilateral deep brain stimulation implants
L. Golestanirad, B. Keil, … L. L. Wald·bioRxiv·Nov 22, 2018
- The effect of device configuration and patient body composition on image artifact and RF heating of deep brain stimulation devices during MRI at 1.5T and 3T
B. Bhusal, B. T. Nguyen, … L. Golestani Rad·bioRxiv·Apr 9, 2020
- RF heating of deep brain stimulation implants in open-bore vertical MRI systems
L. Golestani Rad, D. Lampman, … J. Kirsch·bioRxiv·May 27, 2019
- Effect of field strength on RF power deposition near conductive leads: A simulation study of SAR in DBS lead models during MRI at 1.5 T - 10.5 T
E. Kazemivalipour, A. Sadeghi-Tarakameh, … L. G. Rad·bioRxiv·Dec 13, 2022
- Reducing RF-Induced Heating Near Implanted Leads Through High-Dielectric Capacitive Bleeding of Current (CBLOC)
L. Golestanirad, L. M. Angelone, … L. L. Wald·bioRxiv·Nov 19, 2018
- Artifacts can be deceiving: The actual location of deep brain stimulation electrodes differs from the artifact seen on magnetic resonance images
Noa B Nuzov, Bhumi Bhusal, … Laleh Golestanirad·Stereotactic and Functional NeurosurgeryPeer-reviewed·Dec 16, 2022