RF Heating of Bipolar Epicardial Implants during MRI at 0.55 T and 1.5T: Effect of Device Length and Termination Conditions
Bhumi BhusaliD, Pia SanpitakiD, Fuchang JiangiD, Gregory WebsteriD, Jacob Richardson, Nicole SeiberlichiD, Laleh GolestaniradiD
Abstract
Pediatric patients with epicardial cardiac implantable electronic devices (CIEDs) are frequently excluded from the Magnetic Resonance Imaging (MRI) primarily due to RF heating safety concerns. In this study we evaluate RF heating of two bipolar epicardial leads during MRI at 0.55 T and 1.5 T under different termination conditions. Our findings showed that the mean RF heating was significantly reduced at 0.55 T MRI compared to that at 1.5 T. Similarly, the RF heating at 0.55 T MRI was highest for full system whereas, during MRI at 1.5 T, the RF heating was highest for the capped abandoned lead, showing dependence of RF heating pattern on MRI field strength. While RF heating at both fields surpassed the safety limit, the capped abandoned leads at 1.5 T MRI showed significantly higher RF heating with temperature rise surpassing 50{degrees}C in some of the cases. These results highlight the difference in RF heating of bipolar epicardial leads compared to the previously reported findings for monopolar epicardial lead which showed smallest heating for capped abandoned lead at both field strengths. These findings emphasize the necessity of device-specific evaluations at each field-strength to inform clinical decision-making and expand MRI access for this vulnerable population.
02 /Specification & Classification
Click any tag to browse matching papers- Citations
- 0
- Approach
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Modifying the trajectory of epicardial leads can substantially reduce MRI-induced RF heating in pediatric patients with a cardiac implantable electronic device
F. Jiang, B. Bhusal, … L. Golestanirad·bioRxiv·Jul 26, 2022
- Age and lead configuration matter: A comparative study of RF-induced heating of epicardial and endocardial electronic devices in adult and pediatric anthropomorphic phantoms in 1.5 T MR
- Cumulative Transfer Function for Assessment of MRI-Induced RF Heating Risk in Pediatric Patients Implanted with Bifurcated Leads
Fuchang Jiang, Jasmine Vu, … Laleh Golestanirad·bioRxiv·Jul 10, 2026
- Reducing RF-Induced Heating Near Implanted Leads Through High-Dielectric Capacitive Bleeding of Current (CBLOC)
L. Golestanirad, L. M. Angelone, … L. L. Wald·bioRxiv·Nov 19, 2018
- Effect of field strength on RF power deposition near conductive leads: A simulation study of SAR in DBS lead models during MRI at 1.5 T - 10.5 T
E. Kazemivalipour, A. Sadeghi-Tarakameh, … L. G. Rad·bioRxiv·Dec 13, 2022
- The effect of device configuration and patient body composition on image artifact and RF heating of deep brain stimulation devices during MRI at 1.5T and 3T
B. Bhusal, B. T. Nguyen, … L. Golestani Rad·bioRxiv·Apr 9, 2020