On the Stability of Silicone-Encapsulated CMOS ICs for Active Implantable Devices: 4.3 Years of Accelerated Life Testing
Ahmad Shah Idil, Callum A. Lamont, Kambiz Nanbakhsh, Federico Mazza, Vasiliki Giagka, Timothy G. Constandinou, Anne Vanhoestenberghe, Nick Donaldson
This study evaluates the long-term stability of silicone-encapsulated CMOS ICs in biofluid environments through 4.3 years of accelerated life testing, finding no insulation failures and limited failures at wire bonds.