bioRxivNeurosciencePreprint
Longevity of Implantable Silicon-ICs for Emerging Neural Applications: Evaluation of Bare Die and PDMS-Coated ICs After Accelerated Aging and Implantation Studies
Kambiz NanbakhshiD, Ahmad Shah IdiliD, Callum A. LamontiD, Csaba DücsőiD, Ömer Can AkguniD, Domonkos HorváthiD, Kinga TóthiD, István UlbertiD, Federico MazzaiD, Timothy G. ConstandinouiD, Wouter A. SerdijniD, Anne VanhoestenbergheiD, Nick DonaldsoniD, Vasiliki GiagkaiD
Abstract
Silicon integrated circuits (ICs) are central to the next-generation miniature active neural implants, whether packaged in soft polymers for flexible bioelectronics or implanted as bare die for neural probes. These emerging applications bring the IC closer to the corrosive body environment, raising reliability concerns, particularly for long-term clinical use. Here, we evaluated the long-term electrical and material stability of silicon-ICs from two foundries, after one-year accelerated in vitro and in vivo animal studies. The ICs featured various custom-designed test structures and were partially PDMS coated, creating two regions on each chip, uncoated "bare die" and "PDMS-coated". During the accelerated in vitro study, ICs were electrically biased and periodically monitored. Results demonstrated stable electrical performance for at least a year, suggesting that bare die ICs can function in the body for months. Despite electrical stability, material analysis revealed chemical and electrically driven degradation of the IC passivation in the bare die regions. In contrast, PDMS-coated regions revealed no such degradation, making PDMS a highly suitable encapsulant for ICs intended for years-long implantation. Based on the new insights, guidelines are proposed that may enhance the longevity of implantable ICs, significantly broadening their applications in the biomedical field.
- Citations
- 0
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- On the Stability of Silicone-Encapsulated CMOS ICs for Active Implantable Devices: 4.3 Years of Accelerated Life Testing
Ahmad Shah Idil, Callum A. Lamont, … Nick Donaldson·bioRxiv·May 21, 2025
- Stability of Thin-Film Metallization in Flexible Stimulation Electrodes: Analysis and Improvement of in vivo Performance
P. Cvancara, T. Boretius, … T. Stieglitz·bioRxiv·May 22, 2019
- Ceramic Packaging in Neural Implants
K. Shen, M. M. Maharbiz·bioRxiv·Jun 26, 2020
- On the Reliability of Chronically Implanted Thin-Film Electrodes in Human Arm Nerves for Neuroprosthetic Applications
P. Cvancara, G. Valle, … T. Stieglitz·bioRxiv·May 29, 2019
- Intracortical probe arrays with silicon backbone and microelectrodes on thin polyimide wings enable long-term stable recordings in vivo
Antje Kilias, Yu‐Tao Lee, … Patrick Ruther·Journal of Neural EngineeringPeer-reviewed·Nov 15, 2021
- Planar Amorphous Silicon Carbide Microelectrode Arrays for Chronic Recording in Rat Motor Cortex
J. R. Abbott, E. N. Jeakle, … S. F. Cogan·bioRxiv·Nov 2, 2023