Cryoablation temperature monitoring with dense ultrasonic speed-of-sound shift imaging

Gaya Lamm, Tal GrutmaniD, Mike BismuthiD, Tali IlovitshiD

PDFDOIPreprint

Abstract

Accurate temperature monitoring during cryoablation, a minimally invasive technique that destroys tissue locally by forming an ice ball around an inserted cryoprobe, is vital for achieving complete ablation while protecting surrounding tissue. We present a dense slowness-shift imaging method that estimates local speed-of-sound changes from ultrasound B-mode images using optical flow. This single-transducer, image-based approach enables mapping of spatial temperature change without requiring additional hardware. Cryoablation experiments in a tissue-mimicking phantom and ex vivo turkey breast demonstrated that slowness deviation increases with decreasing temperature. In the phantom, the dependence was linear (a = -20.70 s{middle dot}m-1C{degrees}-1), while in turkey breast it presented an exponential relationship (t = 34.04xexp(0.075(-{Delta}T)) s{middle dot}m-1C{degrees}-1). The algorithm detected sub-degree temperature variations and accurately tracked cooling down to -39.4 {+/-} 5.6 {degrees}C. This work demonstrates the feasibility of ultrasound-based, noninvasive temperature monitoring during cryoablation, providing a scalable, real-time alternative to existing invasive or high-cost thermal assessment techniques.

Citations
0
FWCI
0.00
Field percentile
22th percentile

Reproducibility ledger

Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.

Sign in to file a reproducibility report.