Assessment of thermochromic phantoms for characterizing microwave ablation devices
Ghina ZiaiD, Amber Lintz, Clay Hardin, Anna BottiglieriiD, Jan ŠebekiD, Punit PrakashiD
Abstract
Background and PurposeThermochromic gel phantoms provide a controlled medium for visual assessment of thermal ablation device performance. However, there are limited studies reporting on the comparative assessment of ablation profiles assessed in thermochromic gel phantoms against those in ex vivo tissue. The objective of this study was to compare microwave ablation zones in a thermochromic tissue mimicking gel phantom and ex vivo bovine liver, and to report on measurements of the temperature dependent dielectric and thermal properties of the phantom. MethodsThermochromic polyacrylamide phantoms were fabricated following a previously reported protocol. Phantom samples were heated to temperatures in the range of 20 - 90 {degrees}C in a temperature-controlled water bath, and colorimetric analysis of images of the phantom taken after heating were used to develop a calibration between color changes and temperature to which the phantom was heated. Using a custom, 2.45 GHz water-cooled microwave ablation antenna, ablations were performed in fresh ex vivo liver and phantoms using 65 W applied for 5 min or 10 min (n = 3 samples in each medium for each power/time combination). Broadband (500 MHz - 6 GHz) temperature-dependent dielectric and thermal properties of the phantom were measured over the temperature range 22 - 100 {degrees}C. ResultsColorimetric analysis showed that the sharp change in gel phantom color commences at a temperature of 57 {degrees}C. Short and long axes of the ablation zone in the phantom (as assessed by the 57 {degrees}C isotherm) for 65 W, 5 min ablations were aligned with extents of the ablation zone observed in ex vivo bovine liver. However, for the 65 W, 10 min setting, ablations in the phantom were on average 23.7% smaller in short axis and 7.4 % smaller in long axis than those observed in ex vivo liver. Measurements of the temperature dependent relative permittivity, thermal conductivity, and volumetric heat capacity of the phantom largely followed similar trends to published values for ex vivo liver tissue. ConclusionThermochromic tissue mimicking phantoms provide a controlled, and reproducible medium for comparative assessment of microwave ablation devices and energy delivery settings, though ablation zone size and shapes may not accurately represent ablation sizes and shapes observed in ex vivo liver tissue under similar conditions.
02 /Specification & Classification
Click any tag to browse matching papers- Citations
- 3
- FWCI
- 0.52
- Field percentile
- top 40%
- Approach
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Cryoablation temperature monitoring with dense ultrasonic speed-of-sound shift imaging
Gaya Lamm, Tal Grutman, … Tali Ilovitsh·Physics in Medicine and BiologyPeer-reviewed·May 8, 2026
- Mathematical modeling of the frozen zone dynamics: towards using thermal imagers in cryotherapy
O. V. Ivakhnenko, Oleksandr Todrin, … Sergey N Shevchenko·PLoS ONEPeer-reviewed·Apr 16, 2026
- Study of flow effects on temperature-controlled radio-frequency ablation using phantom experiments and forward simulations
- High Resolution Acoustic Mapping of Gelatin-Based Soft Tissue Phantoms
H. Badawe, P. Raad, M. Khraiche·bioRxiv·May 10, 2023
- Quantification of thermal side effects during pulsed field ablation
Marcela Mercado‐Montoya, Tatiana Gomez Bustamante, … Lawrence Overzet·bioRxiv·Nov 2, 2024
- Dual cryo-radiofrequency ablation enhances lesion depth in beating human left ventricular preparations
Matthew S. Sulkin, Jacob I. Laughner, … Igor R. Efimov·bioRxiv·Nov 14, 2018