bioRxivNeurosciencePreprint
An optimized framework for simultaneous EEG-fMRI at 7T enabling safe, high-quality human brain imaging with millisecond temporal resolution and sub-millimeter spatial resolution
Cristina Sainz MartineziD, Jonathan WirsichiD, Cilia JaegeriD, Tracy WarbrickiD, Serge VulliémoziD, Mathieu LemayiD, Jessica BastiaanseniD, Roland WiestiD, João JorgeiD
Abstract
The combination of electroencephalography (EEG) and functional magnetic resonance imaging (fMRI) at ultra-high field (7 Tesla) offers appealing new possibilities to probe human brain function non-invasively with high coverage, millisecond temporal precision and sub-millimeter spatial precision, unraveling cortical layers and small subcortical structures. Unfortunately, this technique has remained largely inaccessible at 7T, due to prohibitive cross-modal interference effects and physical constraints. Here, we developed a first-of-its-kind EEG-fMRI acquisition framework on a clinical 7T system combining key improvements from previous research: compact EEG transmission to reduce artifact incidence, reference sensors for artifact correction, and adapted leads for compatibility with a dense radiofrequency receive-array allowing state-of-the-art fMRI sensitivity and acceleration. Two implementations were tested: one using an EEG cap adapted in-house, and another using a recently-designed prototype from an industrial manufacturer, intended to be further developed into a commercial device accessible to the broader community. A comprehensive evaluation in humans showed that simultaneous acquisitions, including sub-millimeter fMRI resolution, could be conducted without detectable safety issues or major practical constraints. The EEG exerted relatively mild perturbations on fMRI quality (6-11% loss in temporal SNR), without measurably affecting the detection of resting-state networks and visual responses. The artifacts induced on EEG could be corrected to a degree where the spatial, spectral and temporal characteristics were comparable to outside recordings, and hallmark features such as resting-state alpha and eyes-closing alpha modulation could be clearly detected. Altogether, these findings indicate excellent prospects for neuroimaging applications, that can leverage the unique possibilities achievable at 7T.
- Citations
- 0
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Safety and data quality of EEG recorded simultaneously with multi-band fMRI
Maximillian K. Egan, Ryan J. Larsen, … Sepideh Sadaghiani·PLoS ONEPeer-reviewed·Jul 2, 2021
- Reducing the Risk of Gradient Induced EEG Hardware Heating During Simultaneous EEG-fMRI
Brandon T Ingram, R Allen Waggoner, … Karen J Mullinger·bioRxiv·Nov 3, 2026
- Online Closed-Loop Real-Time tES-fMRI for Brain Modulation: Feasibility, Noise/Safety and Pilot Study
Beni E. Mulyana, Aki Tsuchiyagaito, … Hamed Ekhtiari·bioRxiv·Apr 10, 2021
- The relationship between EEG and fMRI connectomes is reproducible across simultaneous EEG-fMRI studies from 1.5T to 7T
Jonathan Wirsich, João Jorge, … Serge Vulliémoz·NeuroImagePeer-reviewed·Feb 15, 2021
- Automated Electrodes Detection during simultaneous EEG/fMRI
Mathis Fleury, Christian Barillot, … Pierre Maurel·Frontiers in ICTPeer-reviewed·Jan 13, 2019
- Hi-Fi fMRI: High-resolution, fast-sampled and sub-second whole-brain functional MRI at 3T in humans
Benedetta Franceschiello, Simone Rumac, … Eleonora Fornari·bioRxiv·May 13, 2023