Safety and data quality of EEG recorded simultaneously with multi-band fMRI
Maximillian K. EganiD, Ryan J. LarseniD, Jonathan WirsichiD, Bradley P. SuttoniD, Sepideh SadaghianiiD
Abstract
PurposeSimultaneously recorded electroencephalography and functional magnetic resonance imaging (EEG-fMRI) is highly informative yet technically challenging. Until recently, there has been little information about EEG data quality and safety when used with newer multi-band (MB) fMRI sequences. Here, we measure the relative heating of a MB protocol compared with a standard single-band (SB) protocol considered to be safe. We also evaluated EEG quality recorded concurrently with the MB protocol on humans. Materials and MethodsWe compared radiofrequency (RF)-related heating at multiple electrodes and magnetic field magnitude, B1+RMS, of a MB fMRI sequence with whole-brain coverage (TR = 440 ms, MB factor = 4) against a previously recommended, safe SB sequence using a phantom outfitted with a 64-channel EEG cap. Next, 9 human subjects underwent eyes-closed resting state EEG-fMRI using the MB sequence. Additionally, in three of the subjects resting state EEG was recorded also during the SB sequence and in an fMRI-free condition to directly compare EEG data quality across scanning conditions. EEG data quality was assessed by the ability to remove gradient and cardioballistic artifacts along with a clean spectrogram. ResultsThe heating induced by the MB sequence was lower than that of the SB sequence by a factor of 0.73 {+/-} 0.38. This is consistent with an expected heating ratio of 0.64, calculated from the square of the ratio of B1+RMS values of the sequences. In the resting state EEG data, gradient and cardioballistic artifacts were successfully removed using traditional template subtraction. All subjects showed an individual alpha peak in the spectrogram with a posterior topography characteristic of eyes-closed EEG. The success of artifact rejection for the MB sequence was comparable to that in traditional SB sequences. ConclusionsOur study shows that B1+RMS is a useful indication of the relative heating of fMRI protocols. This observation indicates that simultaneous EEG-fMRI recordings using this MB sequence can be safe in terms of RF-related heating, and that EEG data recorded using this sequence is of acceptable quality after traditional artifact removal techniques.
02 /Specification & Classification
Click any tag to browse matching papers- Citations
- 15
- FWCI
- 1.06
- Field percentile
- top 26%
- Organism
- Assay
- Research Focus
- Approach
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Reducing the Risk of Gradient Induced EEG Hardware Heating During Simultaneous EEG-fMRI
Brandon T Ingram, R Allen Waggoner, … Karen J Mullinger·bioRxiv·Nov 3, 2026
- An optimized framework for simultaneous EEG-fMRI at 7T enabling safe, high-quality human brain imaging with millisecond temporal resolution and sub-millimeter spatial resolution
Cristina Sainz Martinez, Jonathan Wirsich, … João Jorge·bioRxiv·Aug 23, 2024
- Data-driven beamforming techniques to attenuate ballistocardiogram (BCG) artefacts in EEG-fMRI without detecting cardiac pulses in electrocardiography (ECG) recordings
Makoto Uji, Nathan E. Cross, … Christophe Grova·Human Brain MappingPeer-reviewed·Jun 8, 2021
- Simultaneous MRI-EEG during a motor imagery neurofeedback task: an open access brain imaging dataset for multi-modal data integration.
Giulia Lioi, Claire Cury, … Christian Barillot·Scientific DataPeer-reviewed·Jun 10, 2020
- The relationship between EEG and fMRI connectomes is reproducible across simultaneous EEG-fMRI studies from 1.5T to 7T
Jonathan Wirsich, João Jorge, … Serge Vulliémoz·NeuroImagePeer-reviewed·Feb 15, 2021
- A multi-session simultaneous EEG-fMRI dataset with repeated experience sampling
Lotus Shareef-Trudeau, David Braun, … Aaron Kucyi·bioRxiv·Feb 7, 2026