Image Quality Causes Substantial Bias in Three-Dimensional Speckle-Tracking Echocardiography Measures
Abstract
BACKGROUNDThree-dimensional speckle-tracking echocardiography (3D-STE) is increasingly used to assess left ventricular (LV) mechanics but the quantitative effect of image quality on measurements is not known.\n\nOBJECTIVESTo assess the impact of image quality on 3D-STE derived LV indices.\n\nMETHODSEchocardiography was performed in two groups of 18 healthy participants. In the first study, optimal and intentionally poor-quality images were acquired. In the second study, a sheet of ultrasound-attenuating material (neoprene rubber) of three different thicknesses (2, 3 and 4 mm) was used to mimic mildly, moderately and severely impaired image quality respectively.\n\nRESULTSIn both studies sub-optimal image quality resulted in a systematic underestimation bias in all LV deformation and rotational indices. LV ejection fraction and volumes were also consistently underestimated. The extent of the bias was proportional to the impairment in image quality (i.e. the poorer the image quality the larger the bias). Reproducibility was also less good for sub-optimal images, although LV volumes and ejection fraction showed excellent reproducibility irrespective of image quality.\n\nCONCLUSIONSSub-optimal image quality introduces a substantial systematic bias and impairs the reproducibility of 3D-STE. Bias related to image quality might have important clinical implications since its magnitude is similar to that reported in association with disease and may confound associations between disease and LV mechanics.\n\nCONDENSED ABSTRACTThree-dimensional speckle-tracking echocardiography (3D-STE) is increasingly used in clinical practice. We assessed the impact of sub-optimal image quality on left ventricular (LV) deformation indices either by poor echocardiographic technique or by impairing ultrasound propagation using a neoprene sheet. Sub-optimal image quality introduces a substantial systematic bias that increases with poorer image quality. Bias due to image quality might have important clinical implications since its magnitude is similar to that reported in association with disease.
02 /Specification & Classification
Click any tag to browse matching papers- Organism
- Assay
- Approach
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Anthropomorphic left ventricular mesh phantom: a framework to investigate the accuracy of SQUEEZ using Coherent Point Drift for the detection of regional wall motion abnormalities
A. Manohar, G. M. Colvert, … E. R. McVeigh·bioRxiv·Nov 13, 2019
- Three-dimensional tissue strain measurement using a row-column array during biaxial testing of excised ventricular myocardium
Xavier Navy, Zhiyu Sheng, … John M. Cormack·Ultrasound in Medicine & BiologyPeer-reviewed·Jun 12, 2025
- In-silico heart model phantom to validate cardiac strain imaging
Tanmay Mukherjee, Muhammad Usman, … Reza Avazmohammadi·Computers in Biology and MedicinePeer-reviewed·Aug 7, 2024
- Echocardiographic evaluation of normal adult left Ventricular geometry in a Nigerian population.
D. c. ugwuanyi, J. c. Eze, … C. U. Eze·bioRxiv·Apr 9, 2020
- Saak Transform-Based Machine Learning for Light-Sheet Imaging of Cardiac Trabeculation
Y. Ding, V. Gudapati, … T. K. Hsiai·bioRxiv·Oct 4, 2019
- High Volume Rate 3-D Ultrasound Imaging Using Fast-Tilting and Redirecting Reflectors
Z. Dong, S. Li, … P. Song·IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency ControlPeer-reviewed·Aug 1, 2023