Vendor-neutral sequences (VENUS) and fully transparent workflows improve inter-vendor reproducibility of quantitative MRI
Abstract
PurposeWe developed an end-to-end workflow that starts with a vendor-neutral acquisition and tested the hypothesis that vendor-neutral sequences decrease inter-vendor variability of T1, MTR and MTsat measurements. MethodsWe developed and deployed a vendor-neutral 3D spoiled gradient-echo (SPGR) sequence on three clinical scanners by two MRI vendors. We then acquired T1 maps on the ISMRM-NIST system phantom, as well as T1, MTR and MTsat maps in three healthy participants. We performed hierarchical shift function analysis in vivo to characterize the differences between scanners when the vendor-neutral sequence is used instead of commercial vendor implementations. Inter-vendor deviations were compared for statistical significance to test the hypothesis. ResultsIn the phantom, the vendor-neutral sequence reduced inter-vendor differences from 8 - 19.4% to 0.2 - 5% with an overall accuracy improvement, reducing ground truth T1 deviations from 7 - 11% to 0.2 - 4%. In vivo we found that the variability between vendors is significantly reduced (p = 0.015) for all maps (T1, MTR and MTsat) using the vendor-neutral sequence. ConclusionWe conclude that vendor-neutral workflows are feasible and compatible with clinical MRI scanners. The significant reduction of inter-vendor variability using vendor-neutral sequences has important implications for qMRI research and for the reliability of multicenter clinical trials.
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Toward Large-Scale Preclinical Neuroimaging: Quantifying Inter-Scanner Variability in Mouse Brain MRI
Moussa Shahid, Jiangyang Zhang·bioRxiv·Jun 2, 2026
- A resource for development and comparison of multi-modal brain 3T MRI harmonisation approaches
Shaun Warrington, Asante Ntata, … Stamatios N. Sotiropoulos·Imaging NeurosciencePeer-reviewed·Nov 20, 2023
- Normalization of MRI T1w between-scan effects for improved longitudinal volumetric estimates
Donatas Sederevičius, Atle Bjørnerud, … Anders M. Fjell·bioRxiv·Jun 19, 2022
- Evaluation of MRI sequences for quantitative T1 brain mapping
P. Tsialios, M. Thrippleton, C. Pernet·bioRxiv·Sep 29, 2017
- A robust intensity distribution alignment for harmonization of T1w intensity values
Donatas Sederevičius, Atle Bjørnerud, … Anders M. Fjell·bioRxiv·Jun 15, 2022
- Increased Brain Volumetric Measurement Precision from Multi-Site 3D T1-weighted 3T Magnetic Resonance Imaging by Correcting Geometric Distortions
N. D. Nanayakkara, S. R. Arnott, … R. Bartha·bioRxiv·Nov 29, 2021