Increased Brain Volumetric Measurement Precision from Multi-Site 3D T1-weighted 3T Magnetic Resonance Imaging by Correcting Geometric Distortions
N. D. Nanayakkara, S. R. Arnott, C. J. M. Scott, I. Solovey, S. Liang, V. S. Fonov, T. Gee, D. N. Broberg, S. M. H. Haddad, J. Ramirez, C. Berezuk, M. Holmes, S. Adamo, M. Ozzoude, A. Theyers, S. Sujanthan, M. Zamyadi, L. Casaubon, D. Dowlatshahi, J. Mandzia, D. Sahlas, G. Saposnik, A. Hassan, R. H. Swartz, S. C. Strother, , R. Bartha
Abstract
Magnetic resonance imaging (MRI) scanner-specific geometric distortions may contribute to scanner induced variability and decrease volumetric measurement precision for multi-site studies. The purpose of this study was to determine whether geometric distortion correction increases the precision of brain volumetric measurements in a multi-site multi-scanner study. Geometric distortion variation was quantified over a one-year period at 10 sites using the distortion fields estimated from monthly 3D T1-weighted MRI geometrical phantom scans. The variability of volume and distance measurements were quantified using synthetic volumes and a standard quantitative MRI (qMRI) phantom. The effects of geometric distortion corrections on MRI derived volumetric measurements of the human brain were assessed in two subjects scanned on each of the 10 MRI scanners and in 150 subjects with cerebrovascaular disease (CVD) acquired across imaging sites. Geometric distortions were found to vary substantially between different MRI scanners but were relatively stable on each scanner over a one-year interval. Geometric distortions varied spatially, increasing in severity with distance from the magnet isocenter. In measurements made with the qMRI phantom, the geometric distortion correction decreased the standard deviation of volumetric assessments by 35% and distance measurements by 42%. The average coefficient of variance decreased by 16% in gray matter and white matter volume estimates in the two subjects scanned on the 10 MRI scanners. Geometric distortion correction using an up-to-date correction field is recommended to increase precision in volumetric measurements made from MRI images.
02 /Specification & Classification
Click any tag to browse matching papers- Organism
- Assay
- Approach
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Normalization of MRI T1w between-scan effects for improved longitudinal volumetric estimates
Donatas Sederevičius, Atle Bjørnerud, … Anders M. Fjell·bioRxiv·Jun 19, 2022
- Estimating the effect of a scanner upgrade on measures of grey matter structure for longitudinal designs
Evelyn Medawar, Ronja Thieleking, … Frauke Beyer·PLoS ONEPeer-reviewed·Oct 5, 2021
- Inter-scanner brain MRI volumetric biases persist even in a harmonized multi-subject study of multiple sclerosis
Kelly A. Clark, Carly M. O’Donnell, … the NAIMS Cooperative·bioRxiv·May 5, 2022
- Toward Large-Scale Preclinical Neuroimaging: Quantifying Inter-Scanner Variability in Mouse Brain MRI
Moussa Shahid, Jiangyang Zhang·bioRxiv·Jun 2, 2026
- Empirical field mapping for gradient nonlinearity correction of multi-site diffusion weighted MRI
Colin B. Hansen, Baxter P. Rogers, … Bennett A. Landman·Magnetic Resonance ImagingPeer-reviewed·Nov 19, 2020
- A resource for development and comparison of multi-modal brain 3T MRI harmonisation approaches
Shaun Warrington, Asante Ntata, … Stamatios N. Sotiropoulos·Imaging NeurosciencePeer-reviewed·Nov 20, 2023