Inter-Instrument Quantification of Fluorescence Read-out Signal Using DNA Origami Calibrator Beads
Anja Bille BohniD, Charlotte Christie PeterseniD, James C. S. WoodiD, Finn Skou PederseniD, Denis SelnihhiniD
Abstract
Comparing experimental results across different instruments and laboratories is challenging due to variations in instrument sensitivity and resolution, particularly for analyzing nano-sized and dimly fluorescing particles like viruses and extracellular vesicles. To address this challenge, we conducted a study using fluorescent DNA calibrator beads to test, calibrate, and compare the sensitivity of five different flow cytometers. Our approach involves DNA beads produced through a bottom-up DNA origami method. These beads contain a defined number of fluorophores (0-220) and exhibit no autofluorescence. Fluorophores are precisely conjugated onto the beads in a controlled array, minimizing dye interactions. This allows triggering on one fluorophore and calibration based on another, facilitating precise and accurate calibration in absolute fluorophore numbers. We tested five flow cytometers--FACSAria III, NovoCyte 3000YBG, NovoCyte Quanteon 4500, LSRFortessa, and CytoFlex S--using Cy5 and FAM DNA calibrator beads. We investigated various instrument parameters, including count rate, trigger and calibration gain, gain amplification linearity, and fluorescence trigger sensitivity. Using fluorescent triggering, we demonstrated the capability to detect dimly fluorescent particles, enabling instrument calibration and fluorescence sensitivity comparison using DNA calibrators. In conclusion, our study offers a robust method for addressing instrument fluorescence channel resolution in absolute fluorophore numbers in the analysis of nano-sized and dimly fluorescent particles.
- Citations
- 0
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Intrinsic variability of fluorescence calibrators impacts the assignment of MESF or ERF values to nanoparticles and extracellular vesicles by flow cytometry
E. Lozano-Andres, T. Van Den Broeck, … G. J. A. Arkesteijn·bioRxiv·Mar 1, 2021
- Fluorescence and light scatter calibration allow comparisons of small particle data in standard units across different flow cytometry platforms and detector settings.
J. A. Welsh, J. C. Jones, V. A. Tang·bioRxiv·Apr 21, 2020
- Characterization of nanoparticles and fluorescent recombinant extracellular vesicles with three high-sensitivity flow cytometers
Estefanía Lozano−Andrés, Ye Tian, … Marca H. M. Wauben·bioRxiv·Feb 19, 2026
- Novel PE and APC Tandems: Additional Near-Infrared Fluorochromes for Use in Spectral Flow Cytometry
Y. Seong, Y. Wu, … F. A. Harding·bioRxiv·Jun 25, 2020
- Dedicated nanoparticle flow cytometry for single extracellular vesicle phenotyping: Performance of the CytoFLEX Nano
Johann Mar Gudbergsson, Mette Galsgaard Malle, … Anja Bille Bohn·bioRxiv·Dec 17, 2025
- FLOW CYTOMETRIC PROCEDURES FOR DEEP CHARACTERIZATION OF NANOPARTICLES
Valentina Tirelli, Felicia Grasso, … Massimo Sanchez·Biology Methods and ProtocolsPeer-reviewed·Jan 1, 2025