bioRxivBiochemistryPreprint
Harmonization of IGF1 immunoassay methods using an LC-MS/MS method and associated normative dataset.
Eef G.W.M. LentjesiD, Mark S. PrattiD, Ido P. Kema, Martijn H. J. R. van FaasseniD, Ruben E.A. MussoniD, Michel J VosiD
- Citations
- 1
Abstract
ObjectiveGeneration and testing of IGF1 reference materials (RM), suitable for the harmonization of immunoassay (IA) and LC-MS/MS methods for the IGF1 determination in blood. In addition, establishment of age related reference intervals for men and women. MethodsIn a split sample study of 42 patients, and 30 healthy volunteers we tested the commutability of four RMs for IGF1, using four commercial IAs and an LC-MS/MS method. A new set of age dependent reference intervals was established using Lifelines biobank samples, based on the IGF1 LC-MS/MS method. ResultsThe four RMs were found to be commutable, except the RM with the lowest concentration measured with the Siemens Immulite method. The value assignment of the RMs was based on the IGF1 LC-MS/MS method, which was calibrated against WHO international standard 02/254. LC-MS/MS results were on average about 0 to 60% lower than those of the immunoassays. Combining the recalculated IGF1 results in patient samples from a former study with the data from healthy volunteers in this study, showed a reduction in the variation of the data points (standard error of estimate) of 42% and 62% respectively. ConclusionCommutable RMs for IGF1 can be made from serum of healthy blood donors. However, it remains necessary to test the commutability of these RMs in IAs that were not included in this study. By harmonizing methods using the four RMs, the same age-related reference intervals can be used.
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Sources of variation and establishment of Russian reference intervals for major hormones and tumor markers
A. Ruzhanskaya, K. Ichihara, … V. Emanuel·bioRxiv·May 26, 2020
- Lysosomal Enzyme activity: Establishing Reference Intervals from Patient Data
Dylan Mordaunt, Samantha Stark, … Michael Patrick Metz·medRxiv·Apr 11, 2024
- Enzyme Immunoassays May Be Inaccurate: Familiar Correlates of Male Testosterone (T) Replicate in Britons when T is Measured by Mass Spectrometry but Do Not Replicate in Americans When T is Measured by Enzyme Immunoassay
Allan Mazur, Soazig Clifton·bioRxiv·Jun 20, 2018
- Quartet metabolite reference materials for assessing inter-laboratory reliability and data integration of metabolomic profiling
Naixin Zhang, Peipei Zhang, … Leming M. Shi·bioRxiv·Nov 1, 2022
- Identification and quantification of human relaxin proteins by immunoaffinity-mass spectrometry
Y. Rais, A. P. Drabovich·bioRxiv·Jan 12, 2024
- Universal Study Design for Instrument Changes in Pharmaceutical Release Analytics
Anne B. Ries, Maximilian N. Merkel, … Frederik T. Weiß·ElectrophoresisPeer-reviewed·Jun 1, 2025