Dose assessment in dental cone beam computed tomography: comparison of optically stimulated luminescence dosimetry with the Monte Carlo method
C. Lee, J. Yoon, S.-S. Han, J. Y. Na, J.-H. Lee, Y. H. Kim, J. J. Hwang
Abstract
The usage and the model variety of CBCT machine has been rapidly increasing, the dose evaluation of individual devices became an important issue. Patient dose from CBCT was assessed with two different methods, optically stimulated luminescence dosimeter (OSLD) measured and monte carlo (MC) simulation, in four different examination modes. Through the measurement process and obtained value, more practical and efficient method in acquiring CBCT effective dose would be suggested. Twenty-five OSLD were calibrated and equipped in human phantom of head and neck organs. This was exposed on 2 CBCT units, CS9300 (Carestream Dental LLC, Atlanta, Georgia) and RAYSCAN + (Ray Co. Ltd, Hwaseong-si, Korea) with 2 different examination modes. Dose recorded in dosimetry was obtained and organ dose as well as an effective dose were obtained in each units of examination modes. Those values were also calculated using MC software, PCXMC (STUK, Helsinki, Finland). The organ doses and effective doses from both methods were compared by each examination mode of individual unit. OSLD measured effective dose value was higher than that obtained with MC method in each examination mode, except dual jaw mode of CS9300. The percent difference of effective dose between the two methods were ranged from 4.0 to 14.3 %. The dose difference between the methods was decreased as the examination FOV decreased. Organ dose values were varied according to the method, while overall trend was similar in both methods. The organs showing high dose were mostly consistent in both methods. In this study, the effective dose obtained by OSLD measurement and MC simulation were compared and both methods were described in detail. Consequently, as relatively efficient and easy-handling method, we carefully suggest MC simulation for further dose evaluation.
02 /Specification & Classification
Click any tag to browse matching papers- Assay
- Approach
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Marker-based watershed transform method for fully automatic mandibular segmentation from low-dose CBCT images
Yi Fan, Richard J Beare, … Christopher L. Adamson·Dentomaxillofacial RadiologyPeer-reviewed·Oct 31, 2018
- Clinical super-resolution computed tomography of bone microstructure: application in musculoskeletal and dental imaging
Santeri J. O. Rytky, Aleksei Tiulpin, … Jaakko L. Niinimäki·Annals of Biomedical EngineeringPeer-reviewed·Feb 15, 2024
- Internal calibration for opportunistic computed tomography muscle density analysis
A. Smith, J. J. Tse, … S. L. Manske·bioRxiv·Aug 5, 2022
- Dosimetric Study of Breast Scar Marker in Different Radiotherapy Dose Calculation Algorithms
A. A. A. Mohammed, S. S. Abdullah, … M. Z. Abdul Aziz·bioRxiv·Jan 30, 2019
- Strong correlation between phantomless and inline phantom-based densitometric calibration of vertebral properties from CT scans of healthy volunteers
Fiona G. Gibson, Zichu Ding, … Stefaan W. Verbruggen·Frontiers in Bioengineering and BiotechnologyPeer-reviewed·May 29, 2026
- Does pixel/voxel-size limit the measurement of distances in CBCT-tomography?
Roland Köberle, Deusdedit Lineu Spavieri, … João Roberto Gonçalves·bioRxiv·Dec 3, 2018