Strong correlation between phantomless and inline phantom-based densitometric calibration of vertebral properties from CT scans of healthy volunteers
Fiona G. Gibson, Zichu Ding, Margaret PaggiosiiD, Catherine HandforthiD, Janet E. Brown, X. Li, Enrico Dall’AraiD, Stefaan W. VerbruggeniD
Abstract
Phantom calibration is currently the gold standard for calibrating CT scans and for calculating material properties of dense tissues for computational models. However, in Oncology departments and low-resource settings, it is not routine to include a calibration phantom within the scanning protocol. Therefore, retrospective scan datasets are challenging to calibrate for biomechanical investigations, precluding detailed measurements of material and mechanical properties. In this study, we compared the results from a phantomless calibration technique, where the density within each scan was independently calibrated based on known tissue densities captured within each scan (e.g. air), with those from a traditional inline phantom calibration. To do so we used scans from a cohort of healthy volunteers from the control arm of a clinical trial dataset (ANTELOPE) in which inline calibration phantoms were included. We found that, when selecting air and the aorta as regions for calibration within individual CT scans, a strong individual-specific correlation existed between bone mineral density measured in the phantomless and phantom calibrations. This indicates that the phantomless calibration method can be a useful and reliable tool for quantifying the densitometric material properties of healthy human vertebrae, and provides the opportunity for further analysis of spinal CT scans in either retrospective datasets or in low-resource clinical settings.
- organism: human
- assay: clinical cohort / trial
- research focus: cell type annotation
- research focus: clustering
- research focus: differential expression
- research focus: gene regulatory network inference
- approach: experimental assay
- Citations
- 1
- FWCI
- 1.72
- Field percentile
- top 15%
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Internal calibration for opportunistic computed tomography muscle density analysis
A. Smith, J. J. Tse, … S. L. Manske·bioRxiv·Aug 5, 2022
- Comparing Linear and Nonlinear Finite Element Models of Vertebral Strength Across the Thoracolumbar Spine: A Benchmark from Density-Calibrated Computed Tomography
Matthias Walle, Bryn E. Matheson, Steven K. Boyd·GigaSciencePeer-reviewed·Jan 1, 2025
- Characterizing meniscal calcifications with photon counting-based dual-energy computed tomography
Eeva Alina Nevanranta, Ville‐Pauli Karjalainen, … Mikko Finnilä·Annals of Biomedical EngineeringPeer-reviewed·Jan 1, 2025
- Raman mineral-to-matrix ratios correlate with weight percentage mineral-to-matrix ratio determined by in-SEM Raman imaging of bone tissue
Guillaume Mabilleau, Dale Boorman, Jorge Diniz·bioRxiv·May 30, 2024
- Quantitative Molecular Imaging of Breast Microcalcification Composition using Photon-Counting Spectral Computed Tomography
T. E. Curtis, R. K. Roeder·bioRxiv·Sep 16, 2020
- Upscaling of In Vivo HR-pQCT Images Enables Accurate Simulations of Human Microstructural Bone Adaptation
Nicholas Ohs, Duncan C. Tourolle né Betts, … Ralph Müller·bioRxiv·May 13, 2020