Pauwels type III fracture, wherein the fracture plane makes an angle of 50{degrees}-70{degrees} with the transverse plane, is the most unstable femoral neck fracture. Although there are various types of internal fixation devices, including Cannulated cancellous screws (CCS), dynamic hip screw (DHS), dynamic hip screw along with anti-rotation screws (DHS+ARS), and a newly introduced femoral neck system (FNS), patients with Pauwels type III fracture often experience implant failure. There was no study investigating the performance of these internal fixators under physiological musculoskeletal loading. The objective of the present study was to compare the biomechanical performance of CCS, DHS, DHS+ARS and FNS under daily physiological activities like normal walking and stair climbing. The unstable Pauwels type III fracture was virtually induced in a composite femur bone model and was fixed with all four internal fixators. Second-order tetrahedral elements were used to develop the FE models. The stresses, strains, and deformations of intact and reconstructed femur and implant components were compared to understand the biomechanical efficacy. FNS showed the highest inferior axial displacement of the femoral head while DHS+ARS exhibited the least inferior axial displacement. Among the four fracture fixation systems, DHS+ARS offered the highest post-fixation stability in terms of fracture gap opening (normal walking: 0.82mm and stair climbing: 0.95mm) and contact sliding, while the performance of single hole locking plated FNS was the worst (normal walking: 2.65mm, stair climbing: 3.72mm). Based on this computational study, DHS+ARS seems to be a better performing system in Pauwels type III fracture fixation.
- Citations
- 2
- FWCI
- 1.90
- Field percentile
- top 13%
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Pelvic Ring Fractures: Does Spinopelvic Alignment Affect Sacral and Lumbopelvic Fixation Stability?
Sudharshan Tripathi, Yogesh Kumaran, … Norihiro Nishida·bioRxiv·Oct 17, 2024
- Biomechanical comparison of plate materials and designs for subcondylar fracture fixation: An in silico assessment
A. Gupta, R. Ghosh, … K. Mukherjee·bioRxiv·Aug 7, 2023
- Biomechanical evaluation of a novel orthogonal angle-stable interlocking nail in a canine femur model
D. Lucena, D. V. F. Farias, … L. G. G. G. Dias·bioRxiv·Feb 26, 2021
- Deep behavioral phenotyping tracks functional recovery following tibia fracture in mice
Jonathan E. Layne, Dustin M. Snapper, … Chelsea S. Bahney·Frontiers in PhysiologyPeer-reviewed·Aug 26, 2025
- Intraoperative biomechanics of lumbar pedicle screw loosening following successful arthrodesis
H. B. Pearson, C. J. Dobbs, … J. D. Boerckel·bioRxiv·Jun 8, 2016
- Assessing the Clinical Utility of Finite Element Analysis Using Post-operative CT-Derived Models: A Material Comparison of Multi-level Spinal Fusion Constructs
Rohan Tewari, Robert D. Johnston, … Ciara M. Murphy·bioRxiv·Aug 6, 2026