Blood flow modeling under LVAD physiology. From global circulation to local hemodynamics

P. J. Blanco, J. Grinstein, C. A. Bulant, R. Torii, C. Bourantas, P. Lemos, H. Garcia-Garcia

Abstract

This document presents the modeling strategy to address the in-silico study of different LVAD patho-physiological scenarios. The proposed approach employs a closed-loop lumped-parameter compartmental representation of the global circulation in the cardiovascular system (CVS). The CVS is coupled to a HeartMate 3 LVAD, and different cardiovascular conditions are proposed by modification of model parameters. Once the simulation for these conditions are performed, the cardiac function is analyzed in detail, and the global circulation model delivers flow rate waveforms which are employed as boundary conditions in a 3D hemodynamic simulation. This local circulation model is built using a patient-specific geometry of the aortic arch, containing 7 inlet/outlet boundaries, namely: LVAD cannula, aortic root, left and right subclavian arteries, left and right common carotid arteries and thoracic aorta. This model is exploited to investigate the impact of global cardiovascular conditions in the local hemodynamic features, particularly the wall shear stress (WSS) in different spatial regions.

PDFDOI

02 /Specification & Classification

Click any tag to browse matching papers

Reproducibility ledger

Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.

Sign in to file a reproducibility report.

Similarity Map
SelectedSame fieldCross-field
THISPeighambari '26Santiago '21Alavi '22Pelagi '24Moulton '22Shi '25
70% semantic matchBioengineering
Seyed Babak Peighambari, Tanmay Mukherjee, Emilio A. Mendiola+6 more·International Journal of Engineering Science·Mar 23, 2026· 2 citations