BioelectromagneticsNeurosciencePeer-reviewed journal article
Brain-implanted conductors amplify radiofrequency fields in rodents: advantages and risks
Mihály VöröslakosiD, Omid YaghmazadehiD, Leeor AloniD, Daniel K. SodicksoniD, György BuzsákiiD
Abstract
Over the past few decades, daily exposure to radiofrequency (RF) fields has been increasing due to the rapid development of wireless and medical imaging technologies. Under extreme circumstances, exposure to very strong RF energy can lead to heating of body tissue, even resulting in tissue injury. The presence of implanted devices, moreover, can amplify RF effects on surrounding tissue. Therefore, it is important to understand the interactions of RF fields with tissue in the presence of implants, in order to establish appropriate wireless safety protocols, and also to extend the benefits of medical imaging to increasing numbers of people with implanted medical devices. This study explored the neurological effects of RF exposure in rodents implanted with neuronal recording electrodes. We exposed freely moving and anesthetized rats and mice to 950 MHz RF energy while monitoring their brain activity, temperature, and behavior. We found that RF exposure could induce fast onset firing of single neurons without heat injury. In addition, brain implants enhanced the effect of RF stimulation resulting in reversible behavioral changes. Using an optical temperature measurement system, we found greater than tenfold increase in brain temperature in the vicinity of the implant. On the one hand, our results underline the importance of careful safety assessment for brain implanted devices, but on the other hand, we also show that metal implants may be used for neurostimulation if brain temperature can be kept within safe limits.
- Citations
- 9
- FWCI
- 1.09
- Field percentile
- top 25%
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Effect of field strength on RF power deposition near conductive leads: A simulation study of SAR in DBS lead models during MRI at 1.5 T - 10.5 T
E. Kazemivalipour, A. Sadeghi-Tarakameh, … L. G. Rad·bioRxiv·Dec 13, 2022
- Non-invasive modulation of brain activity and behavior by transcranial radio frequency stimulation
Omid Yaghmazadeh, Leeor Alon, … György Buzsáki·Brain stimulationPeer-reviewed·Jan 17, 2026
- RF heating of deep brain stimulation implants in open-bore vertical MRI systems
L. Golestani Rad, D. Lampman, … J. Kirsch·bioRxiv·May 27, 2019
- In-vivo Measurement of Radio Frequency Electric Fields in Mice Brain
O. Yaghmazadeh, S. Schoenhardt, … G. Buzsaki·bioRxiv·Aug 16, 2022
- Reducing RF-Induced Heating Near Implanted Leads Through High-Dielectric Capacitive Bleeding of Current (CBLOC)
L. Golestanirad, L. M. Angelone, … L. L. Wald·bioRxiv·Nov 19, 2018
- Application of Surgical Lead Management and Reconfigurable Coil Technology to Reduce RF Heating of DBS Implants during MRI at 3T Under Variant Body Compositions
B. Bhusal, B. Elahi, … L. Golestani Rad·bioRxiv·Jun 25, 2020