bioRxivBioengineeringPreprint
Cumulative Transfer Function for Assessment of MRI-Induced RF Heating Risk in Pediatric Patients Implanted with Bifurcated Leads
Fuchang JiangiD, Jasmine VuiD, Bhumi BhusaliD, Yifeng QianiD, Safa HameediD, Daniel J. KimiD, Gregory WebsteriD, Giorgio BonmassariD, Laleh GolestaniradiD
Abstract
PurposeRF-induced heating remains a major barrier to MRI access for patients with epicardial cardiac implantable electronic devices (CIEDs). Although ISO/TS 10974 Tier-3 transfer function (TF) methods are established for unbranched leads, no analogous framework exists for bifurcated leads, in which branch asymmetry and inter-branch coupling may substantially alter heating. We developed and validated a cumulative transfer function (cTF) framework to address this gap. MethodsFollowing ISO/TS 10974 Tier-3 formalism, we measured, calibrated, and validated cTFs for a commercial 35 cm bipolar epicardial lead at 1.5 T. The framework explicitly accounts for branch-specific response and cross-branch coupling. Validation was performed with 24 canonical lead configurations in a homogeneous phantom and, without recalibration, in a heterogeneous anthropomorphic pediatric phantom with clinically derived trajectories. A single-branch TF approximation served as a comparator. The validated cTF was applied to predict RF heating across adult and pediatric human models at multiple imaging landmarks. ResultsCompared with the single-branch TF approximation, the cTF reduced prediction error by nearly 70% in the primary validation dataset. In secondary validation, the cTF maintained low error across clinically relevant trajectories and imaging landmarks. In human models, the framework revealed marked anatomy- and landmark-dependent variation in predicted heating for the tested 35 cm lead, with low predicted heating in pediatric models and substantially higher heating in selected adult chest and upper abdominal imaging scenarios. ConclusionThe cTF provides a validated framework for RF-heating assessment of bifurcated leads and substantially improves prediction accuracy over single-branch TF approximations that neglect branch coupling.
- Citations
- 0
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- RF Heating of Bipolar Epicardial Implants during MRI at 0.55 T and 1.5T: Effect of Device Length and Termination Conditions
Bhumi Bhusal, Pia Sanpitak, … Laleh Golestanirad·bioRxiv·May 29, 2026
- Age and lead configuration matter: A comparative study of RF-induced heating of epicardial and endocardial electronic devices in adult and pediatric anthropomorphic phantoms in 1.5 T MR
- Modifying the trajectory of epicardial leads can substantially reduce MRI-induced RF heating in pediatric patients with a cardiac implantable electronic device
F. Jiang, B. Bhusal, … L. Golestanirad·bioRxiv·Jul 26, 2022
- Reducing RF-Induced Heating Near Implanted Leads Through High-Dielectric Capacitive Bleeding of Current (CBLOC)
L. Golestanirad, L. M. Angelone, … L. L. Wald·bioRxiv·Nov 19, 2018
- The effect of device configuration and patient body composition on image artifact and RF heating of deep brain stimulation devices during MRI at 1.5T and 3T
B. Bhusal, B. T. Nguyen, … L. Golestani Rad·bioRxiv·Apr 9, 2020
- Application of Surgical Lead Management and Reconfigurable Coil Technology to Reduce RF Heating of DBS Implants during MRI at 3T Under Variant Body Compositions
B. Bhusal, B. Elahi, … L. Golestani Rad·bioRxiv·Jun 25, 2020