The effect of device configuration and patient body composition on image artifact and RF heating of deep brain stimulation devices during MRI at 1.5T and 3T
B. Bhusal, B. T. Nguyen, J. Vu, B. Elahi, J. Rosenow, M. J. Nolt, R. Lopez-Rosado, J. Pilitsis, M. DiMarzio, L. Golestani Rad
Abstract
BACKGROUNDPatients with deep brain stimulation (DBS) implants have limited access to MRI due to safety concerns associated with RF-induced heating. Currently, MRI in these patients is allowed only in 1.5T horizontal scanners and with pulse sequences with reduced power. Nevertheless, off-label use of MRI at 3T is increasingly reported based on limited safety assessments. Here we present results of systematic RF heating measurements for two commercially available DBS systems during MRI at 1.5T and 3T. PURPOSETo assess the effect of imaging landmark, DBS lead configuration, and patient body composition on RF heating of DBS leads during MRI at 1.5 T and 3T. STUDY TYPEPhantom study. POPULATION/SUBJECTS/PHANTOM/SPECIMEN/ANIMAL MODELGel phantoms and cadaver brain. FIELD STRENGTH/SEQUENCE1.5T and 3T, T1-weighted turbo spin echo. ASSESSMENTRF heating was measured at tips of DBS leads implanted in brain-mimicking gel. STATISTICAL TESTSNone. RESULTSWe observed substantial fluctuation in RF heating mainly affected by phantom composition and DBS lead configuration, ranging from 0.14{degrees}C to 23.73{degrees}C at 1.5 T, and from 0.10{degrees}C to 7.39{degrees}C at 3T. The presence of subcutaneous fat substantially altered RF heating at electrode tips (-3.06{degrees}C < {Delta}T < 19.05{degrees}C). Introducing concentric loops in the extracranial portion of the lead at the surgical burr hole reduced RF heating by up to 89% at 1.5T and up to 98% at 3T compared to worst case heating scenarios. DATA CONCLUSIONDevice configuration and patient body composition significantly altered the RF heating of DBS leads during MRI at 1.5T and 3T. Interestingly, certain lead trajectories consistently reduced RF heating and image artifact over different imaging landmarks, RF frequencies, and phantom compositions. Such trajectories could be implemented in patients with minimal disruption to the surgical workflow.
02 /Specification & Classification
Click any tag to browse matching papers- Assay
- Approach
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Application of Surgical Lead Management and Reconfigurable Coil Technology to Reduce RF Heating of DBS Implants during MRI at 3T Under Variant Body Compositions
B. Bhusal, B. Elahi, … L. Golestani Rad·bioRxiv·Jun 25, 2020
- RF heating of deep brain stimulation implants in open-bore vertical MRI systems
L. Golestani Rad, D. Lampman, … J. Kirsch·bioRxiv·May 27, 2019
- Effect of field strength on RF power deposition near conductive leads: A simulation study of SAR in DBS lead models during MRI at 1.5 T - 10.5 T
E. Kazemivalipour, A. Sadeghi-Tarakameh, … L. G. Rad·bioRxiv·Dec 13, 2022
- Reconfigurable MRI coil technology can substantially reduce RF heating at the tips of bilateral deep brain stimulation implants
L. Golestanirad, B. Keil, … L. L. Wald·bioRxiv·Nov 22, 2018
- Artifacts can be deceiving: The actual location of deep brain stimulation electrodes differs from the artifact seen on magnetic resonance images
Noa B Nuzov, Bhumi Bhusal, … Laleh Golestanirad·Stereotactic and Functional NeurosurgeryPeer-reviewed·Dec 16, 2022
- Magnetic Resonance Fingerprinting based Thermometry (MRFT): application to ex vivo imaging near DBS leads
E. Qian, P. Poojar, … S. Geethanath·bioRxiv·Jan 11, 2023