His bundle has shorter chronaxie than adjacent ventricular myocardium: Implications for pacemaker programming.
M. Jastrzebski, P. Moskal, A. Bednarek, G. Kielbasa, P. Vijayaraman, D. Czarnecka
Abstract
BackgroundStrength-duration curves for permanent His bundle (HB) pacing are potentially important for pacemaker programming.\n\nObjectiveWe aimed to calculate strength-duration curve and chronaxie of the His bundle (HB) and of the adjacent right ventricular (RV) working myocardium and to analyze zones of selective HB capture and battery current drain when pacing at different pulse durations (PDs).\n\nMethodsConsecutive patients with permanent HB pacing were studied. The RV and HB capture thresholds were assessed at several PDs. Battery current drain and zones of selective HB capture at PDs of 0.1, 0.2, 0.4 and 1.0 ms were determined.\n\nResultsIn the whole group (n =127) the HB chronaxie was shorter than the RV chronaxie. This difference was driven by patients with selective HB pacing (HB chronaxie of 0.47 vs RV chronaxie of 0.79 ms). Strength-duration curve for HB had lower rheobase and its steep portion started at shorter PDs thus creating wider distance - zone of programmable selective HB pacing - between the HB and RV strength-duration curves at shorter PDs. The battery current drain was lower with pacing at PDs of 0.1 - 0.4 ms vs 1.0 ms. Chronaxie adjusted PDs offered lowest current drain.\n\nConclusionFor the first time the strength-duration curves for permanent selective and non-selective HB pacing were determined. Selective HB capture and battery longevity can be promoted by shorter PDs (0.2 ms). Longer PDs (1.0 ms) offer bigger safety margin for RV capture and may be preferable if simultaneous RV capture during HB pacing is desired.\n\n\n\nO_FIG O_LINKSMALLFIG WIDTH=198 HEIGHT=200 SRC=\"FIGDIR/small/623348v1_ufig1.gif\" ALT=\"Figure 1\">\nView larger version (66K):\norg.highwire.dtl.DTLVardef@9f7f35org.highwire.dtl.DTLVardef@15870e4org.highwire.dtl.DTLVardef@15bbdbdorg.highwire.dtl.DTLVardef@5f434_HPS_FORMAT_FIGEXP M_FIG C_FIG
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Programmed His bundle pacing - a novel maneuver for the diagnosis of His bundle capture
M. Jastrzebski, P. Moskal, … D. Czarnecka·bioRxiv·Nov 16, 2018
- Electrocardiographic characterization of non-selective His bundle pacing. Validation of novel diagnostic criteria.
M. Jastrzebski, P. Moskal, … D. Czarnecka·bioRxiv·May 8, 2019
- Left Bundle Branch Area Pacing Preserves Mechanical Strain and Synchrony Compared to Right Ventricular Apical Pacing in an Acute Paired Preclinical Model
Emmanuel Offei, Kyoichiro Yazaki, … Muhammad S. Khan·Journal of Interventional Cardiac ElectrophysiologyPeer-reviewed·Apr 25, 2026
- Programmed deep septal pacing for the diagnosis of left bundle branch capture.
M. Jastrzebski, P. Moskal, … D. Czarnecka·bioRxiv·Sep 30, 2019
- Antitachycardia Pacing is More Effective When Delivered to the Left Bundle Branch Area Compared to the Right Ventricle in a Pre-clinical Model of Ischemic Ventricular Tachycardia
Emmanuel Offei, Yuki Ishidoya, … Derek James Dosdall·bioRxiv·Aug 25, 2025
- A new perspective on slow pacemaking in brain and heart
Arthur Fyon, Oleksandra Pavlova, … Kévin Jehasse·bioRxiv·Oct 30, 2025